Siemens ULTRAMAT 6 Manual page 96

Hide thumbs Also See for ULTRAMAT 6:
Table of Contents

Advertisement

Available languages

Available languages

3.1.4
Discrete Output Block with ULTRAMAT, OXYMAT and CALOMAT
The input value from the discrete output function block in the AUTO mode is the setpoint which is transferred
as a cyclical use date for triggering AUTOCAL using the PROFIBUS of the optional module.
Rel.
Parameter
Index
9
SP_D
10
OUT_D
19
FSAFE_TIME
12
READBACK_D
20
FSAFE_TYPE
21
FSAFE_VAL_D
24
SIMULATE
33
CHECK_BACK
34
CHECK_BACK_MASK not used
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
Description
Transfer value/digital value in the AUTO mode
Calculated output value of the discrete output function block.
Monitoring time in seconds. The failsafe control monitors the bus
communication, in particular the setpoints SP_D and RCAS_IN.
The readback value contains the current state in the device (actual
value).
Defines the reaction of the device if the used setpoint is still bad after
the monitoring time FSAVE_TIME has expired. One of three reactions
can be set:
0:
The substitute value FSAFE_VAL_D becomes effective.
1:
The last valid set value is still issued. The value has the
"Uncertain, last valid value" status.
Substitute value in failure behavior and when reaction type
FSAFE_TYPE = 0 is set.
Simulation value and status for the actual value, enabling of
simulation.
not used
96