Analog Input Block with ULTRAMAT, OXYMAT and CALOMAT
The measured value and status for the process is provided via the analog input block.
Rel.
Parameter
Index
10
OUT
11
PV_SCALE
12
OUT_SCALE
13
LIN_TYPE
16
PV_FTIME
17
FSAFE_ TYPE
18
FSAFE_VALUE
19
ALARM_HYS
21
HI_HI_LIM
23
HI_LIM
25
LO_LIM
27
LO_LO_LIM
30
HI_HI_ALM
31
HI_ALM
32
LO_ALM
33
LO_LO_ALM
34
SIMULATE
35
OUT_UNIT_TEXT
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
Description
Output value and status. Result of the analog input function block
Input scale. As a rule the measuring range of the transducer block. The
lowest value (0%) is represented by 0, the top value (100%) by 1.
Output scale. Representation of user-specific physical size. The lowest
value (0%) is represented by 0, the top value (100 %) by 1. Indication of
user-specific unit.
Selection of curve shape for characteristic curve transformation.
Filter time constant τ of digital filter 1st order.
Parameterization of safe position. Active only in AUTO mode if the
measured or simulation value has a bad status (BAD) or the current mode
O/S is active.
0: The default value FSAFE_VALUE is used for the safe position. The
status raised to uncertain – substitute set.
1: The output value last valid before the failure will still be used. The
status is raised to uncertain – last usable value. If there was no valid
value before, the initial value will be returned with the uncertain – initial
value status.
2: The value will be used for further calculation despite the bad status.
If this parameter is not implemented, the AI will behave as described in • 1.
Default value for safe position.
Limit value hysteresis. For avoiding an alarm surge or waving of signal
flags. After a limit value has been exceeded the value must have moved
the hysteresis in the opposite direction before the limit value signal is reset.
Upper alarm limit
Upper prewarning limit
Lower prewarning limit
Lower alarm limit
Alarm status parameter for upper alarm limit.
Alarm status parameter for upper prewarning limit.
Alarm status parameter for lower prewarning limit.
Alarm status parameter for lower alarm limit.
Simulation value and status, enabling of simulation.
ASCII text to be stored if the unit "text reference" is indicated in
OUT_SCALE.
93