2.8.3
Installing the ULTRAMAT 23 Optional Module
Note
With the ULTRAMAT 23 only the A5E00034504 module can be installed for non-intrinsically safe
connection to PROFIBUS-PA and the A5E00019145 module for connection to the PROFIBUS-DP.
Disconnect the device from the mains
Unscrew and remove the housing cover
Remove the old optional module if installed
Fasten the module with the M3 screws to the rear panel between connectors
Fasten the module to the housing base with plastic bolt (included in add-on kit).
Connect the connectors to the motherboard
Screw on housing cover
Connect the device to the PROFIBUS
Connect the device to the mains
Rear panel
Optional module
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
Motherboard
Fastening bolt
Connect 26-pin
ribbon cable
Fig. 2.3