7.4. Diagnostic Information
The following tables show the assignment of fault signals/maintenance requests from the devices to the
diagnosis bits in the physical block.
7.4.1
ULTRAMAT 6 Diagnostic Information
Fault (S)
Meaning of fault/maintenance request Diagnostic
Maintenance
request (W)
No.
S1
Parameter memory
S2
Chopper motor fault
S3
Microflow sensor
S4
External fault
S5
Rec. chamber temp.
S6
Field device heating faulty
S8
Pressure transducer signal
S11
Reference gas pressure too low
S12
Power supply
S14
Measured value > characteristic curve
upper value
S15
Calibration cancelled
S16
Measuring gas flow too low
W1
Calibration difference too large
W2
Zero point adjusting reserve < 20%
W3
Signal voltage calibration too small
W4
Set clock
W5
Microflow sensor
W6
Temperature sensor LCD display
W7
Receiver chamber temperature >
70°C
W9
External maintenance request
W10
Autocal check difference
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
Diagnosis bit name
byte/bit
1 / 4
DIA_MEM_CHKSUM
2 / 2
DIA_CONF_INVAL
1 / 1
DIA_HW_MECH
1 / 1
DIA_HW_MECH
1 / 5
DIA_MEASUREMENT
1 / 1
DIA_HW_MECH
1 / 1
DIA_HW_MECH
1 / 0
DIA_HW_ELECTR
2 / 1
DIA_SUPPLY
2 / 1
DIA_SUPPLY
2 / 1
DIA_SUPPLY
1 / 1
DIA_HW_MECH
1 / 5
DIA_MEASUREMENT
1 / 5
DIA_MEASUREMENT
2 / 1
DIA_SUPPLY
2 / 5
DIA_MAINTENANCE
2 / 0
DIA_ZERO_ERR
2 / 5
DIA_MAINTENANCE
2 / 5
DIA_MAINTENANCE
2 / 5
DIA_MAINTENANCE
2 / 5
DIA_MAINTENANCE
1 / 3
DIA_TEMP_ELECTR
2 / 5
DIA_MAINTENANCE
2 / 5
DIA_MAINTENANCE
2 / 5
DIA_MAINTENANCE
2 / 5
DIA_MAINTENANCE