Siemens ULTRAMAT 6 Manual page 117

Hide thumbs Also See for ULTRAMAT 6:
Table of Contents

Advertisement

Available languages

Available languages

7. Appendix
7.1. Operation with SIMATIC PDM
Parameters can be set and measured values monitored using the SIMTIC PDM operating program and a PC
or programming unit.
You will find detailed information on the Internet at
http://www.feldgeraete.de/00/produkte/simatic_pdm/infos.html
7.2. Device Data Base File (GSD)
PROFIBUS DP / PA requires files for cyclic data transmission which you can download from the Internet at
German:
English:
siem809c.gsd
(3 KB) PA independant language
siem809e.gsd
(4 KB) DP independant language
sie809cn.bmp
(3 KB) Bitmap_Device
7.3. Technical Data
PROFIBUS PA
Bus voltage
Current consumption
Baud rate
Electrical isolation
PROFIBUS DP
Baud rates
Electrical isolation
PROFIBUS DP / PROFIBUS PA
C2-connections
ULTRAMAT 6 / OXYMAT 6 / CALOMAT 6 / ULTRAMAT 23 / OXYMAT 61
Schnittstelle PROFIBUS DP / PA Interface
A5E00054148-03
http://www.ad.siemens.de/csi/gsd
http://www.ad.siemens.de/csi_e/gsd
(according to DIN IEC 1158-2)
9 ... 32 V (not Exi)
9... 24 V option Exi
10 mA (typical)
31,25 kBit/s
PROFIBUS against other electronics or housing: 1810 VAC
(according to EN 50170)
9.6, 19.2, 93.75, 187.5, 500 kBit/s, 1.5, 3, 6,12 MBit/s
PROFIBUS against other electronics or housing: 50V=
4 connections to master class 2 are supported.
automatic connection clearing 60 s after communications breakdown.
117

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents