This section identifies the DPS300 operator programmable features which may be
used to modify the DPS300 display, failsafe, relay, and/or mA output operation.

Volume Parameters (P050 to P055)

If Readings proportional to volume are desired, adjust the following parameters
P050 Tank Shape
Enter the shape option that matches the clarifier monitored.
If the shape option selected, requires additional clarifier dimension entry, the
associated parameters (as indicated below) may be scroll accessed.
When Operation is "level" (P001 = 1), sludge volume is calculated. Alternatively,
when Operation is "space" (P001 = 2), remaining volume capacity is calculated.
In the RUN mode, Readings are displayed in percent of ( and mA outputs are
proportional to) maximum volume. to convert Readings to volumetric units, see
Max Volume (P051).
Values:
1 =
conic or pyramidal bottom
2 =
3 =
7ML19981BC01

APPLICATION PARAMETERS

0 = volume calculation not required
flat bottom
or
4 =
or
5 =
A
parabolic bottom
6 =
A
7 =
53
(preset)
spherical bottom (P052)
A
sloped bottom (P052)
or
A
flat end
parabolic ends (P052 /
P053)
A
L
universal linear (P054 /
P055)
9 =
universal curved (P054 /
P055)
10 =