P818 (V) Profile Pointer Distance
Observe the distance (in P005 Units) between the transducer face and the Profile
Pointer.
Values:
P819 (V) Profile Pointer Amplitude
Observe the amplitude (in dB above 1 µV rms) of the Echo Profile at the Profile
Pointer position.
Values:
P820 Algorithm
Use this feature to select the Algorithm(s) (mathematical operations) on which the
Sonic Intelligence™ echo selection is based.
This value is automatically altered by Material (P002).
If the wrong echo is processed, select an alternate algorithm, while observing the
resultant echo processing displays.
To select an Algorithm...
1. Press
2. Press
3. Press
Algorithms are characterized by a combination of echo characteristics:
area
largest = echo with the largest amplitude above the TVT
first
last
Values:
ALF
A
L
F
AL
AF
LF
bLF
bL
bF
LASt
7ML19981BC01
0.000 to 9999
0 to 99
to display the Auxiliary Function symbol,
or
to display the desired Reading display symbol(s),
(when the desired Algorithm is displayed)
= echo with most area above the TVT
= first echo above the TVT to be received
= last echo above the TVT to be received
= Area, Largest, and First average.
= Area only.
= Largest only
= First only
= Area and Largest average
= Area and First average
= Largest and First average
= Largest or First
= Largest only
= First only
= last only (first echo from empty)