Specifications; Electronics - Siemens interranger DPS300 Instructional Manual

Table of Contents

Advertisement

Electronics

Power:
Environmental:
Scan Points:
Accuracy:
Resolution:
Memory:
! EEPROM (non-volatile) no back-up battery required
Programming:
Display:
area
Temperature :
Compensation
Temperature Error:
Outputs:
! transducer drive:
! relays:
! mA output:
* program range is defined as the empty distance to the face of the transducer
(P006) plus any range extension (P801).
7ML19981BC01
! 100/115/200/230 V AC
! location
! altitude
! ambient temperature
! relative humidity
! installation category
! pollution degree:
! 2 points per DPS300
! 1% of program range* or 2 cm (0.8"), whichever is greater
(ideal conditions)
! via removable programmer and / or Dolphin Plus interface package
! custom graphics, backlit LCD with 51 x 127 mm (5 x 2") viewing
! -50 to 150 °C (-58 to 302 °F)
! integral transducer sensor
! 2 TS-3 temperature sensors, max.
! programmable fixed temperature
! with compensation:
! fixed temperature:
! 150 to 315 V peak
! 1 form "C" SPDT contact per relay,
! rated 5 A at 250 V AC, non-inductive
! 2 outputs max.
! 0.1% resolution
! 0-20 or 4-20 mA, scalable
! 750 Ω, isolated, 30 V rms
5

SPECIFICATIONS

±
15%, 50/60 Hz, 31 VA
! indoor / outdoor
! 2000 m max
! - 20 to 50°C (-5 to 122 °F)
! suitable for outdoor (Type 4X /
NEMA 4X / IP65 enclosure)
! II
! 4
! 1% of program range* or 2 cm (0.8"),
whichever is greater (ideal
conditions)
! 0.1% of range
! 0.22% / °C deviation from
programmed temperature.
! 4 alarm/control relays
InterRanger DPS300

Advertisement

Table of Contents
loading

Table of Contents