Adjustment Of The Stand-Up Angle - Invacare Dragon Vertic Service Manual

Hide thumbs Also See for Dragon Vertic:
Table of Contents

Advertisement

SERVICE MANUAL
8.8

Adjustment of the stand-up angle

Pre-requisites:
· Screwdriver, flat, width of edge approx. 6 mm
WARNING: The Verticalizer could be destroyed if the position of the upper shut-off
switches is changed!
· Never modify the position of the upper shut-off switches!
The control box (1) for the limit stop of the verticalizer is to be
found under the seat.
·
Raise the verticalizer.
·
Use the screwdriver to carefully release the rubber cover
at the upper end of the control box and remove.
There are three limit switches in the control box, two at the
upper end and one at the lower end. The lower limit switch (1)
is the one used to adjust the stand-up angle. Never change the
position of the upper shut-off switches!
66
®
Invacare
- Dragon Vertic

Advertisement

Table of Contents
loading

Table of Contents