Checking A Regulating Motor; Replacing The Verticalizer Actuator - Invacare Dragon Vertic Service Manual

Hide thumbs Also See for Dragon Vertic:
Table of Contents

Advertisement

SERVICE MANUAL
7.13

Checking a regulating motor

Parts/tools required:
· Multimeter
·
The electric resistance of the regulating
motor is to be checked at the connection
assembly. In case this is close to infinite, the
motor is probably burnt out. If it is less than
1W, the motor has short-circuited. In either
case the motor must be replaced.
7.14

Replacing the verticalizer actuator

Pre-requisites:
· Screwdriver, flat, width of edge approx. 6 mm
·
Raise the actuator. The actuator retaining pins are
to be found at (1) and (2).
46
®
Invacare
- Dragon Vertic