SERVICE MANUAL
7.13
Checking a regulating motor
Parts/tools required:
· Multimeter
·
The electric resistance of the regulating
motor is to be checked at the connection
assembly. In case this is close to infinite, the
motor is probably burnt out. If it is less than
1W, the motor has short-circuited. In either
case the motor must be replaced.
7.14
Replacing the verticalizer actuator
Pre-requisites:
· Screwdriver, flat, width of edge approx. 6 mm
·
Raise the actuator. The actuator retaining pins are
to be found at (1) and (2).