Invacare
- Dragon Vertic
7.6

Checking and replacing the main fuse

CAUTION! Fire hazard!
· Always use an original strip fuse with approved current.
· If the main fuse has blown, remedy the cause before a new fuse is installed.
WARNING: Danger of fire exists if the fuse is
incorrectly assembled!
· Only ever assemble the fuse in the order shown in the
illustration at right!
· Always tighten nuts to between 3.3 and 3.5 Nm!
Pre-requisites:
· 8mm open-ended spanner
· 40A strip fuse
Information
If the fuse holder is damaged this can be replaced completely together with the battery cables.
·
Remove the battery cover.
·
If your powerchair is equipped with the
Shark Electronic System, then the fuse
holder is located on top of the batteries (1), if
you have ACS Electronics, then it is located
on the device support.
·
Open the fuse holder cover.
·
If the strip fuse has blown, determine the
cause of the fault and remedy if necessary.
The main fuse should only be replaced
when the fault has been remedied.
SERVICE MANUAL
Until 03.2004
After 03.2004
1. Fuse
1. Fuse
2. Fuse compartment
2. Fuse compartment
3. Washer
3. Nut DIN 6923
4. Nut M5