Rem24 Error Codes And Diagnostics Codes - Invacare Dragon Vertic Service Manual

Hide thumbs Also See for Dragon Vertic:
Table of Contents

Advertisement

SERVICE MANUAL
6.1.3

REM24 error codes and diagnostics codes

The drive electronic system can automatically rectify certain disturbances. In this case the status
display stops flashing. Please switch the remote on and off several times. Wait approximately 5
seconds each time before switching the remote on again. In case the error is not rectified by this,
assess the cause on the basis of the following flashing code.
Flashing code:
Flashes once Module faulty
Flashes twice
Flashes three times Fault on r.h. motor.
Flashes four times
Flashes five times
18
Significance:
Faulty accessory (e.g.
short circuit on regulating
motor)
Lifter too high or too low
(seat not at drive height)
Connection loose or faulty
or motor faulty
Fault on l.h. motor.
Connection loose or faulty
or motor faulty
Fault/brake fault on r.h.
motor. Connection loose
or faulty or motor faulty
Invacare
Solution:
Replace faulty module
·
Check
accessory
connections,
check
accessories
·
Whilst the lifter
is raised, lower
slowly until the
status display
stops flashing.
If the lifter is
too low, raise
slowly until the
status display
stops flashing.
Only drive
when the seat
is at the
correct drive
height.
·
Check the
connecting plug
·
Check the motor
·
Check the
connecting plug
·
Check the motor
·
Check the connecting
plus
®
- Dragon Vertic
Flashing code:
See "Replacing
the Shark
Electronics" on
page 28
See "Checking a
regulating motor
" on page 46
See instruction
manual
See "Checking
the cables" on
page 35
See "Replacing a
motor" on page
24
See "Checking
the cables" on
page 35
See "Replacing a
motor" on page
24
See "Checking
the cables" on
page 35
See "Replacing a
motor" on page
24

Advertisement

Table of Contents
loading

Table of Contents