Diagnostics involving disturbances with electric regulating motors

Assess the cause of the disturbance involving an electric regulating motor on the basis of the
following table:
PROBLEM
Electric
regulating
motor not
reacting
OTHER
POSSIBLE
SYMPTOMS
CAUSE
Remote displays
Light
flashing "E",
/regulating
status diode on
module faulty
light/regulating
module does not
extinguish even
if the remote is
shut down or
disconnected
None
Cable possibly
severed or
damaged
Electric
regulating
motor possibly
faulty
Remote
possibly faulty
SOLUTION
DOCUMENTATION
·
See "Replacing the Shark
Replace
light /regulating
Electronics" on page 28
module
·
See "Checking the
Check that
cables" on page 35
the cable has
not been
severed or
damaged.
Replace cable
if necessary
·
See "Checking a
Test
regulating
regulating motor " on
motor
page 46
·
See "Replacing the Shark
Replace
II Remote" on page 37
remote on
wheelchair in
order to
exclude the
possibility that
the remote is
the cause of
the
disturbance.
SERVICE MANUAL
17