Removing the Heat Exchanger Assemblies

When required
Tools required
Parts required
Preparations for this procedure:
Turn off the column compartment.
Disconnect the power cable.
Disconnect capillaries.
Remove column compartment from stack
and place it on the working bench.
Remove the front cover, top cover and top
foam section, see
the Top Cover"
1100 Series TCC Reference Manual
If the heater is leaking, blocked or does not heat/cool or other assemblies have
to be removed
Screwdriver Pozidriv 1 PT3
Wrench 1/4 – 5/16 inch for capillary connections
Heat exchanger assembly (left) G1316-60007, includes ambient temperature
sensor
Heat exchanger assembly (right) G1316-60006
"Installing the Foam and
on page 96.
Repairing the Column Compartment
1 Disconnect the heat exchanger cables from the processor
board and the grounding connection of the column
switching valve.