Troubleshooting and Test Functions

Defective Heater Circuit

The electronic circuit for control of the heater assemblies is defective.
The processor checks the function of the heater circuits continually. If a defect
is detected in the control circuit, the processor switches off the heater
(peltier) assemblies, and the error message is generated.
Probable Causes
• Defective CCM board.
Suggested Actions
42
Exchange the CCM board.