1
Introduction to the Variable Wavelength Detector
Introduction to the Detector
Introduction to the Detector
The Agilent variable wavelength detectors described in this manual are
designed for highest optical performance, GLP compliance and easy
maintenance with:
• higher data rate up to 240 Hz for ultra- fast- HPLC (G7114B),
• deuterium lamp for highest intensity and lowest detection limit over a
• optional flow- cell cartridges (standard 10 mm, 14 µL; high pressure
• Dual wavelength mode, see
• easy front access to lamp and flow cell for fast replacement,
• electronic identification of flow cell and lamp with RFID (Radio
• built- in electronic temperature control (ETC) for improved baseline
• built- in holmium oxide filter for fast wavelength accuracy verification.
10
wavelength range of 190 to 600 nm,
10 mm, 14 µL; micro 3 mm, 2 µL; semi- micro 6 mm, 5 µL) and Prep
Cells are available and can be used depending on the application needs
(other types may be introduced later),
Frequency Identification) tag for unambiguous identification,
•
lamp information: part number, serial number, production date,
ignitions, burn time
•
cell information: part number, serial number, production date,
nominal path length, volume, maximum pressure
stability, and
"Dual- Wavelength Mode"
Agilent 1290 Infinity II VWD User Manual
on page 17.