Performance Test Record For Your Power Supply; Cv Performance Test Record - Agilent Technologies E3647A User Manual

Table of Contents

Advertisement

Appendix Service Information

Performance Test Record for Your Power Supply

Performance Test Record for Your Power Supply

CV Performance Test Record

Test Description
CV Programming Accuracy
@ 0 volts (DVM reading)
CV Readback Accuracy
@ 0 volts
CV Programming Accuracy
@ Full Scale (DVM reading)
CV Readback Accuracy
@ Full Scale
CV Load Effect (Load
Regulation)
CV Source Effect (Line
Regulation)
CV PARD (Normal mode)
Load Transient Response
Time
194
Models
Output
Actual Result
Output1
all
Output2
Output1
all
Output2
Output1
E3646A/48A
Output2
Output1
E3647A/49A
Output2
Output1
E3646A/48A
Output2
Output1
E3647A/49A
Output2
E3646A/48A
all
E3647A/49A
all
E3646A/48A
all
E3647A/49A
all
E3646A/48A
all
E3647A/49A
all
all
all
Specifications
Upper Limit
Lower Limit
+0.0100 V
-0.0100 V
+0.0250 V
+0.0250 V
DVM + 0.0050 V DVM - 0.0050 V
DVM + 0.0250 V DVM - 0.0250 V
20.0200 V
19.9800 V
20.0450 V
19.9550 V
60.0400 V
59.9600 V
60.0850 V
59.9150 V
DVM + 0.0150 V DVM - 0.0150 V
DVM + 0.0450 V DVM - 0.0450 V
DVM + 0.0350 V DVM - 0.0350 V
DVM + 0.0850 V DVM - 0.0850 V
Maximum change: < 5 mV
Maximum change: < 9 mV
Maximum change: < 5 mV
Maximum change: < 9 mV
< 5 mVp-p / 0.5 mVrms
< 8 mVp-p / 1 mVrms
μ
< 50
sec

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

E3646aE3648aE3649a

Table of Contents