2 - Verification and Performance Tests

Performance Test Record Form

Model _________________
Test Description
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Programming
Front Panel Display Readback
High Voltage (Full Scale) Programming
Front Panel Display Readback
Load Effect
Source Effect
PARD (Ripple and Noise)
Peak-to-Peak
RMS
Transient Response
Voltage in 100 µs
Constant Current Tests
Current Programming and Readback
Low current (0A) Programming
Readback Accuracy @ Iout = 20 mA
High Current (Full Scale) Programming
Readback Accuracy @ Iout
Current Sink (@ -1.4A) Readback
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ − 20 mA
PARD (Current Ripple and Noise)
RMS
Load Effect
Source Effect
* Enter your test results in this column
20
Report No _______________ Date __________________
Minimum Specs.
Results*
− 10 mV
_________
_________
Vout − 3 mV
_________
19.980 V
_________
Vout − 9 mV
− 2.0 mV
_________
− 0.5 mV
_________
0 mV
_________
0 mV
_________
0 mV
_________
_________
− 1.0 mA
_________
Iout − 0.25 mA
_________
1.998 A
_________
Iout − 4.3 mA
Isink − 3.65 mA _________ Isink + 3.65 mA
− 2.5 µA
_________
_________
Iout − 22.5 µA
_________
Iout − 22.5 µA
0 mA
_________
_________
− 0.5 mA
_________
− 0.5 mA
Maximum
Measurement
Specs.
Uncertainty
+ 10 mV
1.6 µV
Vout + 3 mV
1.6 µV
20.020 V
335 µV
Vout + 9 mV
335 µV
+ 2.0mV
20 µV
+ 0.5 mV
20 µV
+ 3 mV
872 µV
+ 0.5 mV
50 µV
+ 20 mV
3 mV
+ 1.0 mA
15.2 µA
Iout + 0.25 mA
15.2 µA
2.002 A
252 µA
Iout + 4.3 mA
252 µA
200 µA
+ 2.5 µA
0.1 µA
Iout + 22.5 µA
1.7 µA
Iout + 22.5 µA
1.7 µA
+ 1.0 mA
200 µA
+ 0.5 mA
1.6 µA
+ 0.5 mA
1.6 µA