Performance Test Record Form - Agilent Technologies 6611C Service Manual

System dc power supply
Hide thumbs Also See for 6611C:
Table of Contents

Advertisement

2 - Verification and Performance Tests

Performance Test Record Form

Model Agilent 6611C
Test Description
Constant Voltage Tests
Voltage Programming and Readback
Low Voltage (0V) Vout
Front Panel Display Readback
High Voltage (Full Scale) Vout
Front Panel Display Readback
Load Effect
Source Effect
PARD (Ripple and Noise)
Peak-to-Peak
RMS
Transient Response
Voltage in 100 s
Constant Current Tests
Current Programming and Readback
Low current (0A) Iout
Readback Accuracy (20mA) Iout
High Current (Full Scale) Iout
Readback Accuracy @ Iout
Current Sink Readback
20 mA Range Current Readback
Readback Accuracy @ 0 A
Readback Accuracy @ + 20 mA
Readback Accuracy @ 20 mA
PARD (Current Ripple and Noise)
RMS
Load Effect
Source Effect
22
Report No _______________ Date __________________
Minimum Specs.
5 mV
Vout 2 mV
7.991 V
Vout 4.4mV
2.0mV
0.5mV
N/A
N/A
20 mV
3.32 mA
Iout 0.54 mA
4.9955 A
Iout 10.5mA
Iout 7.1 mA
2.5 A
Iout 22.5 A
Iout 22.5 A
N/A
1 mA
0.5 mA
Results*
Maximum Specs.
__________
+ 5 mV
__________
Vout + 2 mV
__________
8.009 V
__________
Vout + 4.4 mV
__________
+ 2.0mV
__________
+ 0.5 mV
__________
+ 3 mV
__________
+ 0.5 mV
__________
+ 20 mV
__________
+ 3.32 mA
__________
Iout + 0.54 mA
__________
5.0045 A
__________
Iout + 10.5mA
__________
Iout + 7.1 mA
__________
+ 2.5 A
__________
Iout + 22.5 A
__________
Iout + 22.5 A
__________
+ 2.0 mA
__________
+ 1 mA
__________
+ 0.5 mA

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6612c6613c6614c

Table of Contents