Performance Test Record For Agilent E3631A; Cv Performance Test Record - Agilent Technologies E3631A Service Manual

Triple output dc power supply
Hide thumbs Also See for E3631A:
Table of Contents

Advertisement

Chapter 3 Calibration Procedures

Performance Test Record for Agilent E3631A

Performance Test Record for Agilent E3631A

CV Performance Test Record

Test Description
CV Programming Accuracy @ 0 volts
(DVM reading)
+6V supply
+25V supply
-25V supply
CV Readback Accuracy @ 0 volts
+6V supply
+25V supply
-25V supply
CV Programming Accuracy @ Full Scale
(DVM reading)
+6V supply
+25V supply
-25V supply
CV Readback Accuracy @ Full Scale
+6V supply
+25V supply
-25V supply
CV Load Regulation
+6V supply
+25V supply
-25V supply
CV Line Regulation
+6V supply
+25V supply
-25V supply
CV Ripple/Noise
+6V supply
+25V supply
-25V supply
CV Current Noise (Common Mode)
+6V supply
+25V supply
-25V supply
Load Transient Response Time
+6V supply
+25V supply
-25V supply
Specifications
Actual
Result
Upper Limit
+0.0050 V
-0.0050 V
+0.0200 V
-0.0200 V
+0.0200 V
-0.0200 V
DVM + 0.0050 V
DVM - 0.0050 V
DVM + 0.0100 V
DVM - 0.0100 V
-(DVM + 0.0100) V
-(DVM - 0.0100) V
5.9890 V
6.0110 V
24.9675 V
25.0325 V
-25.0325 V
-24.9675 V
DVM + 0.0110 V
DVM - 0.0110 V
DVM + 0.0225 V
DVM - 0.0225 V
-(DVM + 0.0225) V
-(DVM - 0.0225) V
Maximum change:
<2.6 mV
<4.5 mV
<4.5 mV
Maximum change:
<2.6 mV
<4.5 mV
<4.5 mV
<2 mV p-p, 0.35 mV rms for each of the
three outputs
<1.5 μA for each of the three outputs
<50 μsec for each of three outputs
Lower Limit
3
63

Hide quick links:

Advertisement

Table of Contents
loading