Error Messages

Errors are retrieved in first-in-first-out (FIFO) order. The first error returned is the
first error that was stored. Errors are cleared as you read them. When you have read
all errors from the queue, the
The power supply beeps once each time an error is generated.
If more than 20 errors have occurred, the last error stored in the queue (the most recent
error) is replaced with -350, ''Queue overflow''. No additional errors are stored until
you remove errors from the queue. If no errors have occurred when you read the error
queue, the supply responds with +0,
''No error'' over the remote interface or ''NO ERRORS'' from the front panel.
The error queue is cleared by the *CLS (clear status) command or when power is
cycled. The errors are also cleared when you read the queue.
Note that the *RST (reset command) command does not clear the error queue.
Front-panel operation:
Press
Store
(
) key to return to front-panel operation mode if you operate
Local
Local
power supply in remote operation mode.
ERRORS
View
If the
annunciator is on, press
ERROR
scroll through the error numbers. Press
Press
to increase the scrolling speed of the text on the display. All errors are
>
cleared when you exit the menu by pressing the
for about 30 seconds.
Remote interface operation:
SYSTem:ERRor?
Errors have the following format (the error string may contain up to 80 characters).
-102, ''Syntax error''
126
annunciator turns off and the errors are cleared.
ERROR
to view the errors. Use the knob to
View
to view the text of the error message.
>
View
Read and clear one error from the error queue
or let the display time out