Programming The Tests; General Considerations; Programming Parameters; General Measurement Techniques - Agilent Technologies 6571A Service Manual

Gpib dc power supplies series 657xa and 667xa
Table of Contents

Advertisement

Programming the Tests

General Considerations

Procedures are given for programming these tests either from the front panel keypad or from a GPIB controller for a GPIB
controller for 667xA system supplies. The procedures assume you know how to use the front panel keypad or how to
program over the GPIB (see the Power Supply Operating Manual for more information). When using computer-controlled
tests, you may have to consider the relatively slow (compared to computer and system voltmeters) settling times and slew
rates of the power supply. Suitable WAIT statements can be inserted into the test program to give the power supply time to
respond to the test commands.
Hazardous voltages may be present at the power supply output during these tests. They should be
performed only by qualified electronics personnel.

Programming Parameters

Table 2-2 lists the programming voltage and current values for each model. You can enter these values either from the front
panel or from a controller over the GPIB (for system power supplies).
Agilent Model
Agilent 6571A, 6671A
Agilent 6572A, 6672A
Agilent 6573A, 6673A
Agilent 6574A, 6674A
Agilent 6575A, 6675A

General Measurement Techniques

Figure 2-1 shows the setup for most tests. Measure the dc output voltage directly at the sense (+S and -S) terminals.
Connect these terminals for remote sensing (to the +LS and -LS terminals). Be certain to use load leads of sufficient wire
gauge to carry the output current (see Chapter 4 of the Power Supply Operating Manual). To avoid noise pickup, use
coaxial cable or shielded pairs for the test leads. If you use more than one meter or a meter and an oscilloscope, connect
separate leads for each instrument to avoid mutual-coupling effects.

Performance Test Record Sheets

When performing the tests in this chapter, refer to the Performance Test Record sheets supplied at the end of this chapter.
Table 2-6 is for recording common information, such as, the test equipment used and the environmental conditions. Tables
2-7 through 2-11 are dedicated to specific models and contain the acceptable test values and ranges. A place is provided to
record the results of the test.
Table 2-2. Programming Voltage and Current Values
Full Scale
Max. Prog.
Voltage
Voltage
8V
8.190V
20V
20.475V
35V
35.831V
60V
61.425V
120V
122.85V
Full Scale
Max. Prog.
Current
Current
220A
225.23A
100A
102.37A
60A
61.43A
35A
35.83A
18A
18.43A
Max. Prog.
-CC
Overvoltage
Current
10V
10A
24V
10A
42V
5A
72V
4A
144V
2.5A
Verification 13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

6575a6671a6672a6673a6674a6675a ... Show all

Table of Contents