Page 97
1 Select the output to be calibrated and enable the output of the power supply by sending the commands: 2 Disable the voltage protection function. 3 Unsecure the power supply with the secure code before calibration. 4 For voltage calibration, connect a digital voltmeter (DVM) across the power supply’s output terminals.
The Questionable Status Register Table 4-3. Bit Definitions - Questionable Status Register The Questionable Instrument Status Register...
Page 103
The Questionable Instrument Summary Register The Questionable Status Event register is cleared when: The Questionable Status Enable register is cleared when:...
The Status Byte Register Table 4-5. Bit Definitions – Status Byte Summary Register...
Page 106
The Status Byte Summary register is cleared when: The Status Byte Enable register (Request Service) is cleared when: Using Service Request (SRQ) and Serial POLL...
Execution Errors Execution Errors -101 Invalid character -102 Syntax error -103 Invalid separator -104 Data type error -105 GET not allowed -108 Parameter not allowed -109 Missing parameter...
Self-Test Errors Self-Test Errors The following errors indicate failures that may occur during a self-test. Refer to the Service Information for more information.
Calibration Errors Calibration Errors Service Information...
Page 132
Calibration Errors Bad readback cal data Bad OVP cal data Cal OVP status enabled Gain out of range for Gain Error Correction Cal checksum failed, secure state Cal checksum failed, string data Cal checksum failed, store/recall data in location 1 Cal checksum failed, store/recall data in location 2 Cal checksum failed, store/recall data in location 3 Cal checksum failed, DAC cal constants...
Front Panel Voltage and Current Calibration CAL MODE hold down...
Page 196
General Calibration/Adjustment Procedure 5 Save the changes and select the middle voltage calibration point. 6 Enter the reading you obtained from the DVM by using the knob and resolution selection keys. 7 Save the changes and select the high voltage calibration point. 8 Enter the reading you obtained from the DVM by using the knob and resolution selection keys.
Page 197
General Calibration/Adjustment Procedure Notice...
Page 198
General Calibration/Adjustment Procedure ¸ 14 Enter the computed value (DVM reading by shunt resistance) by using the knob and resolution selection keys. 15 Save the change and select the high current calibration point. 8 8 8 8 ¸ 16 Enter the computed value (DVM reading by shunt resistance) by using the knob and resolution selection keys.
Page 199
Calibration Record for Your Power Supply Calibration Record for Your Power Supply...
January 1, 2004 Date Bill Darcy/ Regulations Manager For further information, please contact your local Agilent Technologies sales office, agent or distributor, or Agilent Technologies Deutschland GmbH, Herrenberger Straβe 130, D71034 Böblingen, Germany Revision: B.00.00 Issue Date: Created on 11/24/2003 3:07...