Chapter 8 Specifications

Supplemental Characteristics

Supplemental Characteristics
Table 8-2.
Supplemental Characteristics
Parameter
Output
Low
Programming
Range
Range
High
(maximum
Range
programmable
values)
OVP
Remote Sensing Capability
Voltage drop
Load regulation
Load voltage
Temperature Coefficient, ±(% of output + offset)
Maximum change in output/readback per °C after a 30-minute warm-up
Voltage
Current
Stability, ±(% of output + offset)
Following 30 minutes warm-up, change in output over 8 hours under constant load,
line, and ambient temperature.
Voltage
Current
Output Voltage Overshoot
During turn-on or turn-off of ac power, output plus overshoot will not exceed 1 V if
the output control is set to less than 1 V. If the output control is set to
1 V or higher, there is no overshoot.
E3646A
E3647A
0 to +8.24 V/
0 to +36.05 V/
0 to 3.09 A
0 to 0.824 A
0 to +20.6 V/
0 to +61.8V/
0 to 1.545 A
0 to 0.515 A
1 V to 22 V
1 V to 66 V
Up to 1 V per each lead
Add 5 mV to spec for each 1-volt change in the + output
lead due to load current changes.
Subtract voltage drop in load leads from specified output
voltage rating.
<0.01% + 3 mV (<0.02% + 5mV for output 2)
<0.02% + 3 mA
<0.02% + 2 mV
<0.1% + 1 mA
E3648A
E3649A
0 to +8.24 V/
0 to +36.05 V/
0 to 5.15 A
0 to 1.442 A
0 to +20.6V/
0 to +61.8V/
0 to 2.575 A
0 to 0.824 A
1 V to 22 V
1 V to 66 V
165
8