Adjusting T-Arm Width
NOTE: For this procedure, refer to FIGURE 5.10.
1. Remove the two mounting screws that
secure the arm pad to the arm tube.
2. Turn the arm pad around and
reposition the arm pad on the arm tube.
3. Re‐secure the arm pad to the arm tube
with the two mounting screws. Tighten
securely.
4. Repeat for the opposite side, if
necessary.
Adjusting T-Arm Depth
NOTE: For this procedure, refer to FIGURE 5.11.
1. Remove the two mounting screws that secure the arm pad to the arm tube.
2. Remove the two socket screws that secure the arm tube to the T‐Arm post.
3. Reposition the arm tube on the T‐Arm post:
• Desk Length Arms ‐ to one of three
positions depending on the desired
arm pad depth.
• Full Length Arms ‐ to one of five
positions depending on the desired
arm pad depth.
NOTE: Two additional positions are
obtainable by turning the arm tube 180°.
4. Re‐secure the arm tube to the T‐Arm
post with the two socket screws.
Torque to 60‐70 in.‐lbs.
5. Reattach the arm pad to the arm tube
with the two mounting screws. Tighten
securely.
6. Repeat for the opposite side, if necessary.
Part No 1118364
Arm Pad
Mounting
Screw
FIGURE 5.10 Adjusting T-Arm Width
Arm Pad
Socket
Screws
Mounting
Screw
T-Arm
Post
FIGURE 5.11 Adjusting T-Arm Depth
49
SECTION 5—ARMS
Arm Tube
Mounting
Screw
Arm Tube
Mounting
Screw
Spree XT