Invacare Spectra XTR User Manual
Invacare Spectra XTR User Manual

Invacare Spectra XTR User Manual

Power wheelchair
Hide thumbs Also See for Spectra XTR:
Table of Contents

Advertisement

This manual MUST be given to the user of the product.
BEFORE using this product, read this manual and save for future reference.
Invacare® Spectra XTR Series
Spectra XTR, Spectra XTR² HD
en Power Wheelchair
User Manual

Advertisement

Table of Contents
loading

Summary of Contents for Invacare Spectra XTR

  • Page 1 Invacare® Spectra XTR Series Spectra XTR, Spectra XTR² HD en Power Wheelchair User Manual This manual MUST be given to the user of the product. BEFORE using this product, read this manual and save for future reference.
  • Page 2 All rights reserved. Republication, duplication or modification in whole or in part is prohibited without prior written permission from Invacare. Trademarks are identified by ™and ®. All trademarks are owned by or licensed to Invacare Corporation or its subsidiaries unless otherwise noted.
  • Page 3: Table Of Contents

    4.3 Using the KLICKfix adapter ..... . . 25 Contents 4.4 Adjusting or removing the luggage carrier ... . . 25 5 Adjusting the mobility device to the user's seating 1 General .
  • Page 4 5.6.4 Adjusting the seat width ..... . . 37 5.12.2 Adjusting the depth of the tray / removing the 5.6.5 Adjusting the seat depth (Modulite seat unit) ..37 tray .
  • Page 5 5.16 Standard 80° footrest ......59 6.5.1 Maximum obstacle height ..... . 69 5.16.1 Swivel the footrests outwards and/or remove 6.5.2 Safety information when ascending obstacles .
  • Page 6 9 Transport ........84 9.1 Transport —...
  • Page 7: General

    If you find that the font size in the print version of the user manual Indicates a hazardous situation that could result in is difficult to read, you can download it as a pdf from the Invacare damage to property if it is not avoided.
  • Page 8: Regulations

    European weather conditions. When equipped with an appropriate lighting system, the vehicle is suitable for use on public roads. You should immediately contact an authorized Invacare dealer if the usability of your mobility device 1.5 Indications is restricted due to: The use of this power wheelchair is recommended for the following •...
  • Page 9: Warranty

    General 1.7 Warranty The terms and conditions of the warranty are part of the general terms and conditions particular to the individual countries in which this product is sold. 1.8 Service life We estimate a service life of five years for this product, provided it is used in strict accordance with the intended use as set out in this document and all maintenance and service requirements are met.
  • Page 10: Safety

    Invacare® Spectra XTR Series 2 Safety WARNING! Risk of damage or injury if mobility device is accidentally set into motion 2.1 General safety notes – Switch the mobility device off before you get in, get out or handle unwieldy objects.
  • Page 11 Safety WARNING! CAUTION! Risk of injury when transferring mobility device Risk of injury if maximum permissible load is to a vehicle for transport with the occupant exceeded seated in it – Do not exceed the maximum permissible load (refer – It is always better to transfer the mobility device to a to 11 Technical data, page 95).
  • Page 12: Safety Information On The Electrical System

    Corroded electrical components due to water, liquid – Do not connect any electric devices to your mobility exposure, or incontinent users can result in death, device that are not expressly certified by Invacare for serious injury, or damage. – Minimize exposure of electrical components to water this purpose.
  • Page 13: Safety Information On Electromagnetic Interference

    Safety DANGER! Risk of damage to the wheelchair Risk of death or serious injury A failure in the electric system can lead to unusual Failure to observe these warnings can cause an electrical behavior such as continuous light, no light, or noises short resulting in death, serious injury, or damage to from the magnetic brakes.
  • Page 14: Safety Information On Driving And Freewheel Mode

    Invacare® Spectra XTR Series WARNING! WARNING! Risk of malfunction due to electromagnetic Risk of injury if the mobility device tips over interference – Inclines and declines can only be travelled up to the – Do not switch on or operate portable transceivers or...
  • Page 15 Safety WARNING! WARNING! Risk of injury if the mobility device tips over Risk of breaking down in adverse weather (continued) conditions, i.e. extreme cold, in an isolated area – Never use the mobility device to transport more than – If you are a user with severely limited mobility, we one person.
  • Page 16: Safety Information With Regard To Care And Maintenance

    Such repair the mobility device, elevating legrests must always be and/or service MUST be performed by a qualified lowered during normal travelling. technician. Contact a dealer or Invacare technician. WARNING! CAUTION! Tipping hazard if antitippers are removed,...
  • Page 17: Safety Information Regarding Changes And Modifications To The Mobility Device

    Risk of injuries, and damage to the mobility device, if unapproved backrests are used CAUTION! A retrofitted backrest which is not approved by Invacare Risk of injuries and damage to mobility device due for use with this mobility device may overload the...
  • Page 18: Safety Information On Wheelchairs With A Lifter

    Invacare® Spectra XTR Series Important information about maintenance work CAUTION! tools Risk of injury if the wheelchair tips over – Some maintenance work which is described in this – Never exceed the maximum permissible load (see manual and can be carried out by the user without chapter 11 Technical data, page 95).
  • Page 19: The Position Of The Labels On The Product

    Safety If the mobility device is fitted Important information regarding speed reduction with a tray, it is imperative with raised lifter – If the lifter has been raised above a certain point, the that it is removed and safely drive electronics considerably reduces the speed of stowed when transporting the the wheelchair.
  • Page 20 Invacare® Spectra XTR Series Explanation of symbols on labels This mobility device may be used as a vehicle seat, but Date of manufacture only if it is equipped with a headrest that has been approved by Invacare for this This product complies with Directive 93/42/EEC purpose.
  • Page 21 Safety Do not allow any body parts to get under a raised This symbol indicates the “Drive” position of the seat! coupling lever. In this position the motor is engaged and the motor brakes are operational. You can drive the mobility device. •...
  • Page 22: Components

    3.1 Main parts of the wheelchair Information regarding operation of the lifter at temperatures of less than 0 °C – Invacare mobility aids are fitted with safety mechanisms that prevent capacity overload of the electronic components. At operating temperatures below freezing point this can, in particular, lead to the lifter actuator being shut down after approx.
  • Page 23 Risk of tipping, if the speed limiter sensors fail when the lifter is raised – If you find that the speed reduction function is not working when the lifter is raised, do not drive with the lifter raised and immediately contact an authorized Invacare dealer. 1528224-Q...
  • Page 24: Accessories

    You can find more information about maintenance work on belts in the service manual, which is available from Invacare. 4.2 Using the cane holder If your mobility device is fitted with a cane holder, it can be used for the safe transport of a walking cane, underarm or forearm crutches.
  • Page 25: Using The Klickfix Adapter

    Rixen + Kaul KLICKfix system. To this you can attach various The adapter can turn in 90° steps, allowing you to attach an accessory accessories such as the cellphone case supplied by Invacare, which from any of four different directions. Please refer to the installation you can use to transport your cellphone, sports glasses etc.
  • Page 26 Invacare® Spectra XTR Series Risk of breakage due to excessive load The luggage carrier can break if too great a load is applied. – The maximum permitted load on the luggage carrier is 10 kg. Open the clamp levers (1) of the luggage carrier bracket.
  • Page 27: Adjusting The Mobility Device To The User's Seating Posture

    If the mobility device does not perform to specifications, IMMEDIATELY turn the mobility device Off and reenter set-up specifications. Contact Invacare, if mobility device still does not perform to correct specifications. 1528224-Q...
  • Page 28: Adjustment Possibility For Remote

    Invacare® Spectra XTR Series CAUTION! CAUTION! Damage to mobility device and accident hazard Risk of the remote being pushed backwards It is possible that collisions can occur between mobility during an accidental collision with an obstacle, device components due to various combinations of...
  • Page 29: Adjusting The Height Of The Remote

    Adjusting the mobility device to the user's seating posture 5.3 Adjustment possibilities for the firm seat 5.2.2 Adjusting the height of the remote (only for swing–away remote holders) unit 5.3.1 Setting the height of the armrests Requirements: Requirements: • 1 x 6 mm Allen key •...
  • Page 30: Adjusting The Width Of The Armrests

    Invacare® Spectra XTR Series Loosen clamping lever (1). Adjust armrest to required position. Retighten the clamping lever. Repeat this procedure for the second armrest. 5.4 Adjustment possibilities for the standard Remove the screw (1) using the Phillips screwdriver. seat unit Set the new armrest height by selecting a higher or lower screw hole.
  • Page 31: Adjusting The Width Of The Armrests

    Adjusting the mobility device to the user's seating posture 5.5 Adjustment possibilities for the Flex 2 seat 5.4.2 Adjusting the width of the armrests unit WARNING! Serious injury hazard if one of the armrests 5.5.1 Adjusting width of the Flex 2 standard armrest falls out of its bracket because they have been Requirements: adjusted to a width which exceeds the permissible...
  • Page 32: Adjust The Height Of The Flip-Up Armrest

    Invacare® Spectra XTR Series Release the interior bolts (1) with the Allen key. Set the height for the armrest. Retighten the bolts. Loosen the screws A and move the armrest in depth. Method 2 Tighten the screws securely. 5.5.3 Adjust the height of the flip-up armrest The height of the flip-up armrests can be set using two methods: •...
  • Page 33: Adjusting The Angle For Armrest With The Flip-Up Armrest

    Adjusting the mobility device to the user's seating posture Release the bolts to adjust the armrest angle (1) with the allen key. Release the knob to adjust the armrest angle (1). Set the angle for the armrest. Re-tighten the bolts. Make sure that the Nordlock washers used are re-inserted.
  • Page 34: Set The Resistance Of The Flip-Up Armrest

    Invacare® Spectra XTR Series Remove the rear cover held by the velcro strips. Release the bolts (2) in the torque support (1) using the Allen key. Maintain counter pressure on the nuts (3) with the Allen key. Set the width of the armrest. The possible positions of the To make the armrest easier to move, release the bolt of the armrests are set using the bores in the torque support.
  • Page 35: Setting The Hemi Armrest Rotation

    Adjusting the mobility device to the user's seating posture Put the armrest in the vertical position. Position the armrest horizontally. Release the interior bolts (1) with the Allen key. Re-tighten the bolts. Make sure that the Nordlock washers used are re-inserted. 5.5.8 Setting the Hemi armrest rotation Swivel the armrest inwards or outwards.
  • Page 36: Adjustment Options For The Modulite Seat Unit

    Invacare® Spectra XTR Series 5.6 Adjustment options for the Modulite seat Depending on the side, the screw is accessible from the unit front or the rear. 5.6.1 Adjusting the height of the armrests Requirements: • Allen key 8 mm Loosen the wing screw A.
  • Page 37: Adjusting The Seat Width

    The description of how the width is adjusted is contained in the service instructions for this mobility device. The service instructions can be ordered from Invacare. However, they contain instructions for specially trained service technicians and describe operations that are not intended for the end user.
  • Page 38: Adjusting The Seat Angle

    Invacare® Spectra XTR Series 5.7 Adjusting the seat angle On both sides, loosen the bottom backrest screw A. Do not remove the screws! CAUTION! Move the backrest to the required seat depth. Adjusting the seat tilt or the backrest angle You can adjust the seat depth steplessly.
  • Page 39: Adjusting The Backrest

    Adjusting the mobility device to the user's seating posture 5.8.2 Adjusting the backrest (standard seat) — manually using perforated plate Pull the belt (1) to the rear. The locking pins (2, only the The figure shows the position of the spindle (1) for manual seat angle right-hand side visible in image) are pulled out of the perforated adjustment.
  • Page 40: Adjusting The Backrest (Firm Seat)

    Invacare® Spectra XTR Series 5.8.5 Adjusting the backrest (Flex 2 seat unit) — Remove the screws (1) on both sides of the backrest that hold it in position. manually using the gas pressure spring Set the backrest to the desired angle by selecting a combination of one of the two holes in the backrest frame, and one of the six holes in the metal fixation plate.
  • Page 41: Adjusting The Width Of The Backrest

    For a replacement description, see the service instructions for this mobility device. The service instructions can be ordered from Invacare. However, they contain instructions for specially trained service technicians and describe operations that are not intended for the end user.
  • Page 42: Unit)

    Invacare® Spectra XTR Series 5.8.8 Adjusting the backrest angle (Modulite seat On both sides, loosen and remove the upper backrest screw A. Set the required backrest angle. unit) Use scale B on the backrest for this purpose. Ensure that you CAUTION! set the same angle on both sides.
  • Page 43: Adjusting The Tension Adjustable Backrest

    This can cause the neck to be hyperextended during collisions. – A headrest must be installed. The headrest optionally supplied for this mobility device by Invacare is perfectly suitable for use during transport. – The headrest must be adjusted to the user's ear height.
  • Page 44: Adjusting The Height Of The Headrest Or Neckrest

    Invacare® Spectra XTR Series Requirements: • Allen key 5 mm Loosen the knob A. Adjust the headrest or neckrest to the required height. Retighten the knob. Loosen the screws A , B or the clamping lever C. Adjust the headrest or neckrest to the required position.
  • Page 45: Adjusting The Width

    Adjusting the mobility device to the user's seating posture 5.10.1 Adjusting the width 5.10.3 Adjusting the depth Requirements: • Allen key 5 mm Loosen the knobs A that hold the lateral supports. Adjust the supports to the desired width. Re-tighten the knobs. Loosen the screws A that allows the support pads to slide forwards and backwards.
  • Page 46: Adjusting The Height-Adjustable Push Handle (Flex 2 Seat)

    Invacare® Spectra XTR Series 5.11 Adjusting the height-adjustable push handle (Flex 2 seat) 5.12.1 Laterally adjusting the tray Loosen the wing-screw (1). Adjust the tray towards the left or right. Loosen the knobs (1) that hold the push handle. Re-tighten wing-screw.
  • Page 47: Adjusting The Depth Of The Tray

    Adjusting the mobility device to the user's seating posture 5.13 Vari-F footrest 5.12.2 Adjusting the depth of the tray / removing the tray 5.13.1 Swivelling the footrest/legrest outward and/or removing The small unlocking button is located on the upper section of the footrest/legrest.
  • Page 48: Setting The End Stop Of The Footrest

    Invacare® Spectra XTR Series Set the desired angle. Requirements: Re-tighten the screw. • 1 x 6 mm Allen key 5.13.3 Setting the end stop of the footrest Requirements: • 1 x 6 mm Allen key • 1 x 10 mm open-ended spanner Loosen the screw (1) using the Allen key.
  • Page 49: Adjusting The Length Of The Footrest

    Adjusting the mobility device to the user's seating posture Move the rubber stop to the desired position. Re-tighten the counternut. Use the Allen key to loosen the screw (1) and swivel the footrest upward in order to access the rubber stop. Move the footrest to the desired position.
  • Page 50: Vari-A Legrests

    Invacare® Spectra XTR Series Requirements: • 1 x 5 mm Allen key Press the unlocking button (1) and swivel the footrest/legrest outward. Remove the footrest/legrest in an upward direction. Use the spanner to loosen the screw (1). 5.14.2 Setting the angle Adjust to the desired length.
  • Page 51: Setting The End Stop Of The Legrest

    Adjusting the mobility device to the user's seating posture Loosen the locking knob (1) counter-clockwise at least one turn. Set the desired angle. Turn the knob clockwise to tighten it. 5.14.3 Setting the end stop of the legrest Hit the knob to release the locking mechanism. Requirements: •...
  • Page 52 Invacare® Spectra XTR Series The end position of the legrest is determined by means of a rubber stop (1). Hit the knob to release the locking mechanism. The rubber stop can be screwed in or out A or pushed up or down B.
  • Page 53: Adjusting The Length Of The Legrest

    Adjusting the mobility device to the user's seating posture 5.14.4 Adjusting the length of the legrest CAUTION! Risk of injury due to incorrect adjustment of the footrests and legrests – Before and during every journey it is imperative to ensure that the legrests contact neither the caster wheels nor the ground Use the open-ended spanner to loosen the counternut (1).
  • Page 54: Adjusting The Height Of The Calf Pad

    Invacare® Spectra XTR Series Requirements: • 1 x 10 mm open-ended spanner Use the Allen key to loosen the screws (1). Adjust to the desired position. Re-tighten the screws. Use the open-ended spanner to loosen the nut (1) and remove.
  • Page 55: Unlocking And Swivelling The Calf Pad Backward When Alighting

    Adjusting the mobility device to the user's seating posture 5.14.7 Unlocking and swivelling the calf pad backward 5.14.8 Adjusting the angle–adjustable foot plate when alighting Requirements: • 1 x 5 mm Allen key Press the calf pad straight down. Use the Allen key to loosen both set screws on the foot plate. Adjust to the desired angle.
  • Page 56: Powered Elevating Legrests (Ade Legrests)

    Invacare® Spectra XTR Series Use the Allen key to loosen the set screw on the foot plate (1). Adjust the foot plate to the desired angle or depth. Re-tighten the screw. Press the unlocking button (1) and swivel the legrest outward.
  • Page 57: Adjusting The Length Of The Legrest

    Adjusting the mobility device to the user's seating posture 5.15.3 Adjusting the length of the legrest CAUTION! Risk of injury due to incorrect adjustment of the footrests and legrests – Before and during every journey it is imperative to ensure that the legrests contact neither the caster wheels nor the ground Requirements: •...
  • Page 58: Unlocking And Swivelling The Calf Pad Backward When Alighting

    Invacare® Spectra XTR Series 5.15.6 Unlocking and swivelling the calf pad backward when alighting Press the calf pad straight down. Use the Allen key to loosen the screws (1). Adjust to the desired position. Re-tighten the screws. Unlock the legrest and swivel outward.
  • Page 59: Adjusting The Angle-Adjustable Foot Plate

    Adjusting the mobility device to the user's seating posture 5.15.7 Adjusting the angle–adjustable foot plate Requirements: • 1 x 5 mm Allen key Use the Allen key to loosen the set screw on the foot plate (1). Adjust the foot plate to the desired angle or depth. Use the Allen key to loosen both set screws on the foot plate.
  • Page 60: Adjusting The Length

    Invacare® Spectra XTR Series Loosen the screw (1) using the Allen key, but do not remove Push the release lever inward or outward. The legrest is released. completely. Swivel the legrest inward or outward. Set the legrest to the desired length.
  • Page 61: Setting The Angle

    Adjusting the mobility device to the user's seating posture Press the unlocking lever (1) down. Adjust the legrest to the required angle. Press the unlocking knob (1) and remove the legrests upwards. Release the unlocking lever. The legrest engages. 5.17.2 Setting the angle 5.17.3 Adjusting the length of the legrest CAUTION! CAUTION!
  • Page 62: Adjusting The Depth Of The Calf Pad

    Invacare® Spectra XTR Series Swivel the calf pad to the front. Release bolt (1) with the Allen key and remove. Set the nut on the other side to the required depth. Adjust the calf pad to match the depth of the nut, put the screw Loosen bolt (1) with the Allen key.
  • Page 63: Adjusting The Height Of The Calf Pad

    Adjusting the mobility device to the user's seating posture 5.17.5 Adjusting the height of the calf pad The screws that allow width adjustment of side-mounted legrests are located under the seat (1). Loosen the screws using the open-end spanner. Adjust the legrest to the desired position. Re-tighten the screws.
  • Page 64: Adjusting The Length Of The Legrest - Standard / Junior / Mini Legrests (Firm Seat)

    Invacare® Spectra XTR Series Loosen the grub screw (1), using the Allen key. Release the wing screw (2). Remove the plastic cover cap from the adjustment screw (1). Set the position of the legrest support bracket to the desired Loosen the screw, using the open-ended spanner 13 mm.
  • Page 65 Adjusting the mobility device to the user's seating posture On the underside of the legrest hanger there is a slit A in the outer tube, which allows visual inspection. When adjusting the angle, the inner tube D should only ever be pulled out so far that its end is flush with the edge B Loosen the fixation screws (1) of the pivotal points of the angel of the clamp C.
  • Page 66: Adjusting The Length Of The Legrest - Angle

    Invacare® Spectra XTR Series 5.19.4 Adjusting the length of the legrest — Angle adjustable legrest CAUTION! Risk of injury due to incorrect adjustment of the footrests and legrests – Before and during every journey it is imperative to ensure that the legrests contact neither the caster...
  • Page 67: Usage

    Usage • The rear mirror (if installed) is adjusted so you can look behind 6 Usage at all times without having to bend forward or shift your seating position. 6.1 Driving 6.3 Parking and stationary The maximum load capacity that is stated in the technical data only states that the system is designed for this mass When parking your vehicle or if your vehicle is stationary for a in total.
  • Page 68: Getting In And Out Of The Mobility Device

    Invacare® Spectra XTR Series Disengaging the parking brake device on the side that the remote is on, then depending on the remote version you may be able to detach the remote cable. Pull lever (1) backwards. 6.4 Getting in and out of the mobility device –...
  • Page 69: Taking Obstacles

    Usage 6.5.2 Safety information when ascending obstacles If you do not have sufficient muscle strength, you should ask other persons for help. Use a sliding board, if possible. CAUTION! Risk of tipping over Getting into the mobility device: – Never approach obstacles at an angle but at 90 Position your mobility device as close as possible to your seat.
  • Page 70: Driving Up And Down Gradients

    Invacare® Spectra XTR Series Approach the obstacle or the curb slowly, head-on and at a CAUTION! right angle. Risk of tipping over Depending on the wheel drive type, stop in one of the following – Only ever drive downhill at a maximum of 2/3 of the positions: top speed.
  • Page 71: Use On Public Roads

    Depending on the type of your remote, one of the following will be fitted with an appropriate lighting system. occur when the drive lock-out feature has been activated: Contact your Invacare dealer if you have any questions. • A minus symbol displays on the remote •...
  • Page 72: Disengaging Motors

    Invacare® Spectra XTR Series 6.9.1 Disengaging motors Disengaging the motor Switch off remote. CAUTION! Push the engaging lever down (1). Risk of the mobility device running away The motor is disengaged. – When the motors are disengaged (for push operation...
  • Page 73: Electrical System

    A defective main fuse may be replaced only after checking • Try to provide a 24 hour charge once a week to make sure that the entire electric system. An Invacare specialised dealer both batteries are fully charged. must perform the replacement. You can find information on •...
  • Page 74: How To Charge The Batteries

    WARNING! be left unattended during charging. All charging devices which Risk of electric shock and damage to the batteries are supplied by Invacare comply with these requirements. – NEVER attempt to recharge the batteries by attaching • You cannot overcharge the batteries when using the charger cables directly to the battery terminals.
  • Page 75: Instructions On Using The Batteries

    Electrical system • • Do not leave the batteries in a low state of charge for an When only one red LED is flashing, the Battery Safe feature extended length of time. Charge a discharged battery as soon is enabled. From this time, speed and acceleration is reduced as possible.
  • Page 76: General Instructions On Handling The Batteries

    Invacare® Spectra XTR Series • Goods Rail Transport / Air Transport Ordinances. Batteries may be Only ever transport damaged batteries in an appropriate transported without restrictions, whether by road, rail or by air. acid-resistant receptacle. • Individual transport companies have, however, guidelines which can Wash all objects that have come into contact with acid with possibly restrict or forbid certain transport procedures.
  • Page 77: Maintenance

    Maintenance found in the service manual for this device, which can be obtained 8 Maintenance from Invacare. That manual, however, is intended to be used by trained and authorized service technicians, and describes tasks which 8.1 Maintenance introduction are not intended to be performed by the user.
  • Page 78: Weekly

    Invacare® Spectra XTR Series 8.3.2 Weekly If inspection is not Item Inspection check passed If inspection is not Tires Check that the tires Contact your dealer. Item Inspection check passed (puncture- are undamaged. Armrests/side Check that armrests Tighten the screw or...
  • Page 79: Inspections Performed

    It is confirmed by stamp and signature that all jobs listed in the inspection schedule of the service and repair instructions have been properly performed. The list of the inspection jobs to be performed can be found in the service manual which is available through Invacare.
  • Page 80 Invacare® Spectra XTR Series Stamp of authorized dealer / Date / Signature Stamp of authorized dealer / Date / Signature 4th Annual Inspection 5th Annual Inspection Stamp of authorized dealer / Date / Signature Stamp of authorized dealer / Date / Signature...
  • Page 81: Repair Instructions

    Removing the wheel please see 11 Technical data, page 95, or consult the service manual, available from Invacare (in this connection please see the addresses and phone numbers at the end of this user manual). In case you require assistance, please contact your Invacare dealer.
  • Page 82: Repairing A Flat Tire At The Front

    Invacare® Spectra XTR Series Inflate the tyre a little. CAUTION! Place the cylinder head screws in the wheel rim and tighten the Risk of explosion screws with 10 Nm. The wheel explodes if the air pressure has not been Ensure that the inner tube is not clamped between the wheel released from the wheel before the wheel rim is rim halves.
  • Page 83 Maintenance Remove the tire with inner tube from the wheel rim halves. Repair inner tube and re-fit it, or insert new. If the old inner tube has been repaired and is to be used again, and became wet during repair, it is easier to replace it if it is lightly dusted with talcum powder beforehand.
  • Page 84: Transport

    Invacare® Spectra XTR Series 9.2 Transferring the mobility device to a vehicle 9 Transport WARNING! 9.1 Transport — General information The mobility device is in danger of tipping over if it is transferred to a vehicle while the driver is...
  • Page 85: Use Of The Wheelchair As A Seat In A Vehicle

    Transport 9.3 Use of the wheelchair as a seat in a vehicle CAUTION! There is a risk of injury if the wheelchair is not The following section does not apply to models or properly secured during use as a vehicle seat configurations which may not be used as a vehicle seat.
  • Page 86: How The Wheelchair Is Anchored In A Vehicle

    (UK for example), but may also The power wheelchair is fitted with four anchoring be obtained from Invacare as an option in other countries. points, which are labelled with the symbol shown on the right. Snap hooks or belt loops can be used This mobility device complies with the requirements of ISO for fixation.
  • Page 87: How The User Is Secured Within The Wheelchair

    This can cause the neck to be hyperextended during collisions. – A headrest must be installed. The headrest optionally supplied for this mobility device by Invacare is perfectly suitable for use during transport. – The headrest must be adjusted to the user's ear height.
  • Page 88 Invacare® Spectra XTR Series The pelvic belt should be positioned in the area between the user's pelvis and thighs so that it is unobstructed and not too loose. The ideal angle of the pelvic belt to the horizontal is between 45° and 75°.
  • Page 89: Transporting The Mobility Device Without Occupant

    Invacare strongly recommends that you additionally disconnect or remove the batteries. Refer to Removing the batteries. • Invacare strongly recommends securing the mobility device to the floor of the transporting vehicle. 9.4.1 Preparing the wheelchair for transport without occupant...
  • Page 90: Folding The Backrest Forwards (Standard Seat)

    Invacare® Spectra XTR Series 9.4.3 Folding the backrest forwards (standard seat) Pull the belt (1) to the rear. Pull the battery cases out using the side handles. The locking pins (2, only the right-hand side visible in image) are pulled out of the perforated plates. The backrest can now be moved.
  • Page 91: Removing The Seat (Not Possible For Versions With Lifter)

    Transport 9.4.4 Removing the seat (not possible for versions with lifter) Risk of damage to wheelchair if the seat cannot be folded up or forwards without collision – Remove legrests first. – Ensure that the legrest attachments do not hinder the seat when folding up.
  • Page 92: Special Information About Reassembly

    Invacare® Spectra XTR Series 9.4.6 Special information about reassembly Guide the remote cable under the seat support and connect it to the electronics system. When reassembling, ensure that the seat guide forks (1) are Fold the seat down. It must audibly engage.
  • Page 93 Transport If the backrest has an electrical recliner, refasten the motor to To do this, first place the front of the battery case on the guide the backpost. rods and then push the assembly in. Once the wheelchair is reassembled, you must ensure CAUTION! that the locking system on the rear battery case has Injury hazard if the clevis pin of the motor of...
  • Page 94: After Use

    Invacare® Spectra XTR Series 10 After Use 10.1 Refurbishment The product is suitable for refurbishment. Actions to be carried out: • Cleaning and disinfection. Refer to 8 Maintenance, page 77. • Inspection according to service plan. Consult service instructions, available from Invacare.
  • Page 95: Technical Data

    Technical data 11 Technical data 11.1 Technical specifications The technical information provided hereafter applies to a standard configuration or represents maximum achievable values. These can change if accessories are added. The precise changes to these values are detailed in the sections for the respective accessories. Permissible operating and storage conditions •...
  • Page 96 Invacare® Spectra XTR Series Charging device • Input voltage 200 – 250 V nominal • Operating temperature (surroundings) -25° ... +50 °C • Storage temperature -40° ... +65 °C Drive wheel tires • Tire type 317 (12½" x 2¼") pneumatic or puncture-proof Tire pressure The recommended maximum tire pressure in bar or kpa is marked on the side wall of the tire or the rim.
  • Page 97 Technical data Driving characteristics • with lifter 6° (10.5 %) according to manufacturer’s specifications with 130 kg payload, 4° seat angle, 20° backrest angle • Max. climbable obstacle height 60 mm (100 mm with curb climber) • Turning diameter 1600 mm (without lifter) •...
  • Page 98 Invacare® Spectra XTR Series Seat type Dimensions according to ISO 7176–15 Standard Firm seat Flex 2 / Max • • • Seat height 450 mm (without 460 mm (without lifter) 470 mm (without lifter) • • lifter) 480 - 780 mm (with lifter) 490 - 790 mm (with lifter) •...
  • Page 99 Technical data Seat type Dimensions according to ISO 7176–15 Modulite • • Total height (without lifter) 1100 mm 930 mm (single-part seat frame) • 960 - 1060 mm (telescopic seat frame, moving the backrest plate) • Total height (with lifter) 930 mm (single-part seat frame) •...
  • Page 100 Invacare® Spectra XTR Series Seat type Dimensions according to ISO 7176–15 Modulite • • Backrest height 500 - 620 mm 480/540 mm (sling back) • • 560 - 675 mm 560 - 660 mm (telescopic seat frame, moving the backrest plate) •...
  • Page 101 Technical data Footrests and legrests Seat type Type Firm seat Flex 2 / Max Modulite Standard • Vari F Length 290 - 460 mm • Angle 0 - 70° • Vari A Length 290 - 460 mm • Angle 0 - 70° •...
  • Page 102 Invacare® Spectra XTR Series Payload Standard Firm seat Flex 2 / Max / Modulite • • • • Max. payload 130 kg 60 kg (seat depth 30) 130 kg 75 kg • • 65 kg (seat depth 33) 160 kg (Heavy Duty) •...
  • Page 103 As soon as the lifter is raised above a certain point you can adjust the seat and backrest angles only at a combined total of max. 15°. The actual curb weight depends on the fittings your mobility device has been supplied with. Every Invacare mobility device is weighed when leaving the works.
  • Page 104 Invacare Sales Companies Australia: Ireland: New Zealand: Canada: Invacare Australia PTY. Ltd. Invacare Corporation Invacare Ireland Ltd, Invacare New Zealand Ltd 1 Lenton Place, North Rocks N.S.W. 570 Matheson Blvd E Unit 8 Unit 5 Seatown Business Campus 4 Westfield Place, Mt Wellington,...

This manual is also suitable for:

Spectra xtr² hd

Table of Contents