Use proper cleaning procedures to clean all fiber
connectors before every use. Neglecting this step or
using improper procedures can cause unreliable test
results and may permanently damage the
connectors. See Chapter 4.
Use a video probe to periodically inspect the OTDR
connectors for scratches and other damage.
Read the instructions for splice machines before
using the OTDR to monitor splicing procedures. The
OTDR can interfere with the light injection detection
techniques used by some splicers.

AC Adapter and Battery

You can use the AC adapter (model VERSIV-ACUN) or the lithium
ion battery (model VERSIV-BATTERY) to supply power to the
tester.
To remove the battery, see "Remove the Battery" on page 239.

Charge the Battery

Before you use the battery for the first time, charge the battery
for about 2 hours with the tester turned off.
To charge the battery
Connect the AC adapter to the 15V jack on the left side of the
tester. The LED near the AC adapter connector is red when the
battery charges, and green when the battery is fully charged.
A fully-charged battery operates for approximately 8 hours of
typical use. The battery takes approximately 4 hours to fully
charge when the tester is turned off.