Grating and Filter Motor Test
This test moves the grating motor to the end position, and displays the
motor-position step difference. This checks the movement of the grating drive
mechanism.
The test is available in
• Agilent LabAdvisior (preferred)
• Agilent ChemStation
When to do the Test
• if recalibration cannot be done successfully, or
• if the holmium test fails.
Filter/Grating Motor Test with Agilent LabAdvisor
Figure 44
Agilent 1260 Infinity Variable Wavelength Detector User Manual
Grating and Filter Motor Test with Agilent LabAdvisor
8
Test Functions
131