Probable cause
1
2
Grating or Filter Motor Errors
Error ID: Grating: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809;
Filter: 7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
Test 0 Failed:
Test 1 Failed:
During the motor tests, the detector moves the motor to the end position while
monitoring the end-position sensor. If the end position is not found, the error
message is generated.
Probable cause
1
2
3
4
1220 Infinity II LC System User Manual
Filter motor is not connected.
Cable/connector defective.
Motor is not connected.
Defective motor.
Defective/missing grating or filter.
Cable/connector defective.
Error Information
VWD Detector Error Messages
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Filter motor.
Grating motor.
Suggested actions
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
Please contact your Agilent service
representative.
10
231