Capacitance - Fluke 5502E Getting Started

Multi-product calibrator
Table of Contents

Advertisement

Capacitance

Absolute Uncertainty, tcal ±5 °C
±(% of output + floor)
Range
90 Day
220.0 to 399.9 pF 0.38 + 0.01 nF
0.4 to 1.0999 nF
0.38 + 0.01 nF
1.1 to 3.2999 nF
0.38 + 0.01 nF
3.3 to 10.999 nF
0.19 + 0.01 nF
11 to 32.999 nF
0.19 + 0.1 nF
33 to 109.99 nF
0.19 + 0.1 nF
110 to 329.99 nF
0.19 + 0.3 nF
0.33 to
0.19 + 1 nF
1.0999 μF
1.1 to 3.2999 μF
0.19 + 3 nF
3.3 to 10.999 μF
0.19 + 10 nF
11 to 32.999 μF
0.30 + 30 nF
33 to 109.99 μF
0.34 + 100 nF
110 to 329.99 μF
0.34 + 300 nF
0.33 to
0.34 + 1 μF
1.0999 mF
0.34 + 3 μF
1.1 to 3.2999 mF
3.3 to 10.999 mF
0.34 + 10 μF
11 to 32.999 mF
0.7 + 30 μF
33 to 110.00 mF
1.0 + 100 μF
[1]
The output is continuously variable from 220 pF to 110 mF.
[2]
Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. The maximum allowable peak voltage is
3 V. The maximum allowable peak current is 150 mA, with an rms limitation of 30 mA below 1.1 μF and 100 mA for 1.1 μF and
above.
The maximum lead resistance for no additional error in 2-wire COMP mode is 10 Ω.
[3]
[1] [2] [3]
Resolution
1 Year
0.5 + 0.01 nF
0.1 pF
0.5 + 0.01 nF
0.1 pF
0.5 + 0.01 nF
0.1 pF
0.25 + 0.01 nF
1 pF
0.25 + 0.1 nF
1 pF
0.25 + 0.1 nF
10 pF
0.25 + 0.3 nF
10 pF
0.25 + 1 nF
100 pF
0.25 + 3 nF
100 pF
0.25 + 10 nF
1 nF
0.40 + 30 nF
1 nF
0.45 + 100 nF
10 nF
0.45 + 300 nF
10 nF
0.45 + 1 μF
100 nF
0.45 + 3 μF
100 nF
0.45 + 10 μF
1 μF
0.75 + 30 μF
1 μF
1.1 + 100 μF
10 μF
Allowed Frequency or Charge-Discharge Rate
Min and Max to
Typical Max for
Meet
<0.5 % Error
Specification
10 Hz to 10 kHz
20 kHz
10 Hz to 10 kHz
30 kHz
10 Hz to 3 kHz
30 kHz
10 Hz to 1 kHz
20 kHz
10 Hz to 1 kHz
8 kHz
10 Hz to 1 kHz
4 kHz
10 Hz to 1 kHz
2.5 kHz
10 to 600 Hz
1.5 kHz
10 to 300 Hz
800 Hz
10 to 150 Hz
450 Hz
10 to 120 Hz
250 Hz
10 to 80 Hz
150 Hz
0 to 50 Hz
80 Hz
0 to 20 Hz
45 Hz
0 to 6 Hz
30 Hz
0 to 2 Hz
15 Hz
0 to 0.6 Hz
7.5 Hz
0 to 0.2 Hz
3 Hz
Multi-Product Calibrator
Detailed Specifications
Typical Max for
<1 % Error
40 kHz
50 kHz
50 kHz
25 kHz
10 kHz
6 kHz
3.5 kHz
2 kHz
1 kHz
650 Hz
350 Hz
200 Hz
120 Hz
65 Hz
40 Hz
20 Hz
10 Hz
5 Hz
19