Performance Tests - Fluke 5502E Getting Started

Multi-product calibrator
Table of Contents

Advertisement

5502E
Getting Started

Performance Tests

To make sure that the Product is in specification, use Tables 4 through 12. The tables are
for approved metrology personnel who have access to a standards laboratory that has the
correct equipment to test calibration equipment of this level of accuracy. The tables show
the recommended test points and the permitted maximum and lower limits for each point.
The limits were calculated by adding or subtracting the 90-day specification from the
output value. There is no built-in factor for measurement uncertainty. These performance
tests replace those found in the 5502A Operators Manual and only apply to the 5502E.
Range
329.9999 mV
329.9999 mV
329.9999 mV
3.299999 V
3.299999 V
3.299999 V
3.299999 V
3.299999 V
32.99999 V
32.99999 V
32.99999 V
32.99999 V
32.99999 V
329.9999 V
329.9999 V
329.9999 V
329.9999 V
1000.000 V
1000.000 V
1000.000 V
1000.000 V
1000.000 V
1000.000 V
26
Table 4. Verification Tests for DC Voltage (Normal)
Output
0.0000 mV
329.0000 mV
-329.0000 mV
0.000000 V
1.000000 V
-1.000000 V
3.290000 V
-3.290000 V
0.00000 V
10.00000 V
-10.00000 V
32.90000 V
-32.90000 V
50.0000 V
329.0000 V
-50.0000 V
-329.0000 V
334.000 V
900.000 V
1020.000 V
-334.000 V
-900.000 V
-1020.000 V
Lower Limit
-0.0030 mV
328.9805 mV
-329.0194 mV
-0.000005 V
0.9999855 V
-1.000045 V
3.2899863 V
-3.290136 V
-0.00005 V
9.99955 V
-10.00045 V
32.89863 V
32.90136 V
49.9972 V
328.9846 V
-50.0027 V
-329.0153 V
333.983 V
899.958 V
1019.952 V
-334.016V
-900.042 V
-1020.047 V
Upper Limit
0.0030 mV
329.0194 mV
-328.9805 mV
0.000005 V
1.000045 V
-0.999955 V
3.290136 V
-3.2898638 V
0.00005 V
10.00045 V
-9.99955 V
-32.90136 V
-32.89863 V
50.0027 V
329.0153 V
-49.9972 V
-328.9846 V
334.016 V
900.042 V
1020.047 V
-333.983 V
-899.958 V
-1019.952 V