Airflow Considerations

For safe operation and maintenance of the Product, make sure
that the space around the Product meets minimum
requirements.
Baffles put cool air from the fan into the chassis to internally dissipate heat when the
Calibrator is in operation. The accuracy and dependability of all internal parts of the
calibrator are enhanced by a cool internal temperature. You can increase the life of the
Calibrator and enhance its performance by these rules:
The area around the air filter must be a minimum 3 inches from close walls or rack
enclosures.
The exhaust perforations on the sides of the Calibrator must be clear of blockage.
The air that gets to the Calibrator must be at room temperature. Make sure the
exhaust air from other instruments does not enter into the fan inlet.
Clean the air filter after 30 days or more frequently if the calibrator is operated in a
dusty environment. (See the "Maintenance" chapter of the Operators Manual for
instructions to clean the air filter.)

Instruction Manuals

The 5502E Manual Set includes:
5502A Operators Manual on included CD-ROM (PN 4155227)
5502E Getting Started (PN 4238530)
One of each manual above comes with the instrument. Refer to the Fluke Calibration
Catalog or speak to a Fluke Calibration sales representative (see "Contact Fluke
Calibration") if more printed copies are necessary. The manuals are also available on the
Fluke Calibration website.

5502E Getting Started Manual

This 5502E Getting Started contains a brief introduction to the 5502E Manual Set,
instructions on how to get your calibrator prepared for operation and a complete set of
specifications.

5502A Operators Manual

The 5502A Operators Manual supplies complete information to install the Calibrator and
operate it from the front-panel keys and in remote configurations. The manual also has a
glossary of calibration, specifications, and error-code information. The Operators Manual
includes:
Installation
Controls and features, front-panel operation
Remote operation (IEEE-488 bus or serial port remote control)
Serial port operation (print, show, or transfer data, and setup for serial port remote
control)
Operator maintenance, with verification and calibration procedures
Accessories
SC600 and SC300 oscilloscope calibration options
 Warning
Multi-Product Calibrator
Airflow Considerations
11