Battery - A&D HC-30Ki Instruction Manual

Hc-i series
Hide thumbs Also See for HC-30Ki:
Table of Contents

Advertisement

Using the OP-02 SLA Battery
The scale can be operated with an SLA (Sealed Lead Acid) battery that will be
commercially available.
The scale (with no other options) can be operated for around 80 hours with a fully
charged battery.
The battery will take about 15 hours to be fully charged.
The battery life will vary depending on how the scale is used, the ambient
temperature and so on.
Battery NP4-6
M3 screws
Battery Cover
1. Disconnect the AC adapter from the scale.
2. Loosen the two M3 screws and remove the battery cover.
3. Connect the wires in the battery box to the battery.
Be sure to connect RED wire to positive (+ / RED) terminal and BLACK wire to
negative (- / BLACK) terminal. Or there is a risk of explosion.
4. Place the battery into the box and attach the battery cover using the screws
removed in the step 2 above.
5. Press the ON/OFF key and check that the scale works normally.
12. OP-02 BATTERY
Use a Yuasa Battery NP4-6 (6V, 4Ah).
Use only the AC adapter that is provided with the HC-i scale.
There will be risk of explosion if the battery is connected
improperly or replaced with the incorrect type.
Dispose of a used battery according to the local laws and
regulations.
"-" electrode (BLACK)
"+" electrode (RED)
41