F-06-X RS-232C Data Output
F-06-X requires OP-03 or OP-04 RS-232C interface.
f-06-01
f-06-02
f-06-03
f-06-04
Data Out Mode
Key Mode: Data is sent by pressing the PRINT key.
0
3
+ Command Mode.
Stream Mode: Data is sent continuously.
1
Command Mode cannot be used.
Auto-Print Mode A: Data is sent if the weight display is stable
2
at +5d (weighing display division) and above.
+ Command Mode.
Auto-Print Mode B: Data is sent if the weight display is
3
stable, at ±5d (weighing display division) and above/below.
+ Command Mode
4
Command Mode Only.
5
This setting cannot be used.
6
UFC format with Key Mode (see setting "0").
7
UFC format with Auto-Print Mode A (see setting "2").
8
UFC format with Auto-Print Mode B (see setting "3").
Data to be Sent.
0100
3 Count data sent.
Unit
You select which data to be sent by keying in a
ID
Weight
0 or
weight or unit weight.
0000
Example: Key in 1
this setting would send only the ID number and
Weight
the count.
PCS(count)
Data Format
When f-06-01="6", "7" or "6" is set, the data is UFC format.
Format for AD-8121 MODE 1. Same as f-06-03="2" but the
0
3
interval between continuous data is approximately 2 seconds.
Format for AD-8121 MODE 3. The interval between continuous
1
data is approximately 2 seconds.
Format for general apparatuses, computers, etc.
2
The data-update rate is approximately 10 times per second
when used with the Stream Mode.
Baud Rate
0
3 2400 bps.
1
4800 bps.
2
9600 bps.
1 for the data: ID no., PCS (count),
1
35
0 to display 1100,
0

Hide quick links: