A&D HC-30Ki Instruction Manual page 35

Hc-i series
Hide thumbs Also See for HC-30Ki:
Table of Contents

Advertisement

F-03-X M+ Accumulation Function
f-03-01
f-03-02
+5d
M+
Not Accepted
f-03-02=0
F-04-X Environment and Beeper
f-04-01
f-04-02
f-04-03
f-04-04
M+ Accumulation – Automatic or manual
0
3 Manual accumulation (by pressing the M+ key).
1
Automatic accumulation (Positive data only).
+ or – Count Data Acceptable. (Manual Accumulation)
0
3 Positive data only (5d and above).
1
Positive and negative data (5d and above or –5d or below).
+5d
+
0
-
-5d
M+
Not Accepted
f-03-02=1
Zero Tracking.
Zero tracking traces a drift from zero caused by temperature
change etc., and stabilizes the zero point.
0
3 Zero tracking ON.
1
Zero tracking OFF.
Response
0
Fast / sensitive
1
3 Normal
2
Slow / stable
3
Slower / stabler
Stable Detection Speed / Environment
0
Fast stable detection (good environment).
1
3 Normal.
2
Slow stable detection (poor environment).
Beeper for key operation
0
3 Beeper ON.
1
Beeper OFF.
+