8-2. Gravity Compensation

When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight.
But if the calibration weight cannot be prepared, the gravity acceleration correction will
compensate the scale. Change the gravity acceleration value of the scale to the value
of the area where it will be used. Refer to the gravity acceleration map appended to
the end of this manual.
1. In the Step 1 above, press the TARE key
The display flashes the gravity acceleration
value stored in the scale.
2. Use the 0
display
the
acceleration value.
(Example of the value 9.800 m/s
3. Press the ENTER key.
The scale will store the new value.
If necessary to calibrate the scale
using a weight, go to Step 2 of the
previous section.
4. Press the ON/OFF key to turn the scale off and re-attach the calibration switch
cover. (End of the calibration procedure.)