9-2.

Gravity Compensation

When the scale is first used or has been moved to different place, it should be
calibrated using a calibration weight.
But if the calibration weight cannot be prepared, the gravity acceleration correction
will compensate the scale. Change the gravity acceleration value of the scale to the
value of the area where it will be used. Refer to the gravity acceleration map
appended to the end of this manual.
1. In step 1 of the previous procedure, press
the TARE key.
The display flashes the gravity acceleration
value stored in the scale.
2. Use the 0
display the desired gravity acceleration value.
(Example of the value 9.800 m/s
3. Press the ENTER key.
The scale will store the new value.
If necessary to calibrate the scale
using a weight, go to step 2 of the
previous section.
4. Press the CAL switch and re-attach the
calibration switch cover.
(End of the calibration procedure.)
9 and . 10-key pad to
2
.)
8 8Cal10
.
G r a v i t y A c c e l e r a t i o n
Cal45678g0123aaaaaaaaaa9 . 7 98
M e m o r y G r a v i t y A c c e l e r a t i o n
1234567890123
.
8 8Cal10
G r a v i t y A c c e l e r a t i o n
Cal45678g0123aaaaaaaaa19 . 8 00
[ T A R E ] G r a v i t y A c c e l e r a t i o n
1234567890123
8 8Cal10
.
[ Z E R O ] C a l i b r a t i o n
1234567890123
ENTER
[ T A R E ] G r a v i t y A c c e l e r a t i o n
1234567890123
33
881
pcs
881
pcs
ABC
DEF
GHI
JKL
.
1
2
3
4
5
MNO
PQRS
TUV
WXYZ
#
6
7
8
9
0
C
881
pcs

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents