10 TECHNICAL DATA

The documentation includes a description of the rated characteristics of the electrical connection, all the intake and
outlet connections and the ambient conditions for which the appliance was designed. For full details, also refer to the
"INSTALLATION" section (ref. point 5.4.2 of IEC61010-1:2001).
For "non electric" intakes, the pressure and flow ranges are provided (ref. point 5.4.2 IEC61010-2-040:2005).
Type of water required
Pressure
Type of threaded connection
Required flow [min - max]
Mains water hardness [max]
Fe2+ / Fe3+ [max]
Cold water temperature [max]
Demineralised water conductivity [max]
Limits depending on microbial
contamination
Standard External [Height x Width x
Depth]
Standard External - aquastop accessory
[Height x Width x Depth] - overall height is
increased by +7mm
Standard External - PAD1 accessory
[Height x Width x Depth] - overall depth is
increased by +110mm
Standard External - peristaltic pump P3
accessory [Height x Width x Depth] -
overall depth is increased by +60mm
Net weight
Weight of appliance + packaging
Washing chamber
External surface
User Manual
WATER SUPPLY REQUIRED
1 - Mains (required)
200 kPa – 500 kPa (2 bar – 5 bar)
(for models equipped with a steam condenser, the use of purified water with hardness
Minimum microbiological quality required: "potable water" (ref. limits
in accordance with Italian Legislative Decree 31/2001)
DIMENSIONS and WEIGHT
MATERIALS
GW1160 - GW4060 Series
2 - Demineralised
(recommended)
3 / 4 "
2 - 12 litres / min
42 °f
below 15°f is recommended).
0.5 ppm
35°C
30μS/cm
850 x 600 x 600
857 x 600 x 600
850 x 600 x 710
850 x 600 x 660
72 kg
90 kg
AISI 316L
AISI 304