1-4 Self-Test Mode

Set the switch SWA1 on the bottom of machine to the LOCK position.When power is off,
hold
, and press
NET/GROSS
"Self-Test Mode".
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
Press ON/ZERO key
INTERNAL VALUE MODE
 Press TARE to enter, and the display shows internal value
 Please check whether the internal value has changed obviously with weight
changing.
 Please check the backlight.
 Press ON/ZERO key to back to the last screen , the display shows
KEYPAD TEST MODE
 Press TARE to enter, display shows
Keypad's internal code: TARE key= 07 UNITS key= 06 NET/GROSS key=05
 Press ON/ZERO key to back to the last screen , the display shows
FIRMWARE VERSION DISPLAY MODE
 Press TARE to enter , display shows the firmware version
 Press TARE key again, the display shows maintenance number
 Press ON/ZERO key to back to the last screen, display shows
BACK TO THE LAST SCREEN
Press TARE key to exit self-test mode, the scale will restart automatically.
02005318
icon will be on.
are indicated, the scale is still operational but the
key, Wait till display shows
ON/ZERO
Press UNITS key
Press UNITS key
Press UNITS key
Press UNITS key
(must hook up Load Cell to test)
6
to enter
for 2 secs
ZSME300000298

Hide quick links:

Advertisement

Table of Contents
loading

Table of Contents