7.5 Hold function (animal weighing function)
The balance has an integrated animal weighing function (mean value calculation).
This allows the exact weighing of pets or small animals (load min 1% of max) whilst
the aforesaid are not perfectly still on the weighing platform.
Note: Movement that is too lively (strong display variations) makes a
mean value calculation impossible.
Turn on balance by pressing the [ON/OFF] key. The balance will carry out a
self-test Your balance is ready to weigh as soon as the "0.00 g" display ap-
pears.
Press the [HOLD]-key. The triangle starts flashing on the display whilst the
scales gather several measuring values during this time. Afterwards the calcu-
lated mean value is shown.
To reset the weighing scales to normal weighing mode, press the [HOLD]-key
several times.
This function may be repeated at will by pressing the [HOLD]-key again.
7.6 Plus/minus weighing
For example unit weight control, fabrication control etc.
Switch-on the balance using the [ON/OFF] key and wait for the „0" display.
Place target value on weighing platform and tare it to "0" by pressing the [
TARE ]-key. Remove the nominal weight.
Put the test objects subsequently on the weighing plate, the respective devia-
tion from the nominal weight is displayed with the respective sign to „+" and „-
".
According to the same procedure also packages with the same weight can be
produced, referring to a nominal weight.
Back to weighing mode by pressing the [TARE] button.
22
EOB_M-BA-e-0710