KERN EOB 6K2M Operating Instructions Manual page 13

Platform scale eob_m
Table of Contents

Advertisement

Press the [TARE] key.
The superimposed triangle
top corner. If not, press the [ TARE ]-key.
Press the [HOLD]-key repeatedly until "CAL
0"appears.
Press the [ TARE ]-key, a numeric value is shown on
the display.
After that press the [→0←]-key.
Press the [HOLD]-key
Press the [TARE] key.
Enter the required size of the adjustment weight (See
chpt. 1, "specifications"): To select the digit to be
changed, press the [HOLD]-key and to select the nu-
meric value, press the
[ TARE ]-key.
Confirm by pressing the [→0←]-key.
Place the adjustment weight carefully in the centre of
the weighing platform; the display shows a numeric
value. Press the [HOLD]-key. The adjustment process
is started.
If the adjustment was successful the weighing scales
will return automatically to weighing mode and will
show the weighing value of the adjustment weight.
Take away adjustment weight.
Notice:
For calibrated scales turn off the scales and reset the adjustment switch to calibrat-
ing position.
EOB_M-BA-e-0710
must be at the right hand
CAL U
CAL O
30770
CAL O
CAL 5
u000
0
82077
u000
13

Advertisement

Table of Contents
loading

This manual is also suitable for:

Eob 15k5mEob 60k20mEob 30k10mEob 150k50m

Table of Contents