Fitting the bulb
Filters
Starting up
24
Transmitted-light stand, brightiield, 20 W
&,
Follow the safety instructions given on page 7.
~Fit
the side -faced
col~mn
to the transm itted-l
ig
ht base as
described on page
23.
&
Never touch the bulb when it is hot.
Y
.
b
ou may sustam urns.
• Do nottouch the new 6
V/20
W halogen bulb directly with
your fingers, because finger marks can caus
e
it to crack
when heated.
Remove finger marks immediately w ith a cloth dipped in pure
alcohol.
~
Insert the slider together with t he
bulb socket.
~
Hold the new bulb in a cloth an
d
push
it into position.
~Push
the slider
caref~lly
but complete-
ly back along
t
he ra il to the stop.
• The transmitted-light stand conta ins a
KG1
f ilter.
Yo~
can insert other
0
50
mm filters if you need th
em:
~
Pull out the slider.
~
Insert the
0
50
mm filter.
~
Push the slid er
caref~lly
but
complete ly back along
t
he
rail to the stop
.
The ex-factory setting of the voltage se lector is
230
!for volt-
ages between
200-240
V)
and
t
he fuse holder contains two
160
mAT fuses.
For
100-
120
V power
supply:
~Set
the voltage se lector to
115.
~
Fit
the enclosed two
315
mAt fuse
s
in the f use holder instead ofthe
two
160
mAT
f
uses !see the section
"Changing
fuses"
on,
the right).
~
Connect the powe
r
cable t
o
the
socket
ofthe base and to
t
he power
source.
Lei ca M St€reomicro
sc
op
es
-
Ass
embl
y
Changing fuses
Transmitted or
incident light
Switching on the
transformer
Transmitted light
(bright field), or in-
clined transmitted light
Transmitted light,
bright field
Inclined
transmitted light
Use
:
-
Two
160
mAT fuses for voltages between
200-
240
V.
-
Two
315
mAT fu ses for voltages between
100-
120
V
!see the section "Starting
up"
on the
left).
~
-,11~~'
~
~ ~-·
~Disconnect
the power cable from the stand.
~
~
Pull out the
f~se
holder.
~
Fuse at
rear:
Press
o~t
downwards.
~Spare
fuse at front Pull out sideways and
insert into holder at rear.
~ P~sh
the fuse holder back into po sition.
• Connect an incident lamp to the transmitted-light stand
,
if required
.
• Either transmitted light or incident light can be switched on.
~
For transmitted light
Switch in position
I.
~
For incident light
Switch in position 11.
~
Use thumb wheel to
adjust brightness.
~On
the rear of the stand: Use the thumb
wheel to switch on the transformer and
to adjust the
brightness.
~
Switch on transmitted light.
Bright field is
s~ itable
for examining translucent objects featur-
ing contrasting structures. The obje ct is directly illuminated
fro
m
below and is seen in its natural colo rs against a bright
backgro~nd.
~
For transmitted light, bright field
:
Pu ll lever forwa rds.
Inclined transm itte
d
light whi ch
t
raver ses the object oblique ly
will
p
rod~
c
e
effe cts
advant
ageous for
obs
erving small translu-
cent objects such as foram inifera and fis
h eggs.
~
Move the lever slowly towards the col
um
n of the stand until
th
e
desired effec
t
1s obtain ed.
Leica
VI St ereomi crosco
pes
-
Assembly
25

Hide quick links:

Advertisement

Table of Contents
loading