Floor Stand; Standards; Environmental Conditions; Limitations On Use - Leica M500 OH3 Service Manual

Hide thumbs Also See for M500 OH3:
Table of Contents

Advertisement

Technical data

Floor stand

Model:
Use:
Brakes:
Balancing:
Handgrips:
Stand control unit:
Load:
Space requirement:
Weight:
Range:
44
floor stand with fully-automatic
balancing system and six
electromagnetic permanent
magnetic brakes
carries and positions the Leica
M500 N surgical microscope
six electromagnetic brakes; actua-
tion via pushbutton on handgrip
automatic balancing in all axes and
directions
zoom adjustment
working distance adjustment for
multifocal objective
ALL-FREE button releases all six
brakes on stand
side selector button configurable
with three user-defined brakes
motorised XY adjustment or
navigation of DI C500 functions via
joystick
main/backup illuminator
triggering automatic balancing
status display of balancing
manual balancing of D axis
setting XY speed
switchover FOCUS LOCK - XYZ
FREE
min. 8.0 kg and max. 11.7 kg from
dovetail microscope interface
Foot: 720 x 720 mm
min. height in parked position:
1850 mm
approx. 266 kg with full load
Cantilever: max. 1520 mm
Leica M500 OH3 / Ref. 10 711 894 / Version -

Standards

EN 60601-1
EN 60601-1-2
UL 2601-1

Environmental conditions

Use:
Storage:

Limitations on use

The Leica M500 OH3 is only for use in closed rooms and on a
strong floor.
+10 °C to +40 °C
+50 °F to +104 °F
30 % to 95 % relative humidity
500 mbar to 1060 mbar air pressure
–40 °C to +70 °C
–40 °F to +158 °F
10 % to 100 % relative humidity
500 mbar to 1060 mbar air pressure