3.4.5

Default Configuration

To reset the clamp-on meter to its default parameters (factory configuration):
1.
In the OFF position, hold the
switch to
The "rSt" symbol is displayed.
2.
After 2 s, the clamp-on meter emits a double beep, then all of the digital
symbols of the screen are displayed until the
released. The default parameters are then restored:
Continuity detection threshold = 40
True InRush triggering threshold = 10%
Temperature measurement unit = °C
Adapter function scale factor = 10
3.5

VOLTAGE MEASUREMENT (V)

To measure voltage, proceed as follows:
1.
Set the switch to
2.
Connect the black lead to the COM terminal and the red lead to the "+"
terminal.
3.
Connect the test probes or the alligator clips to the circuit to be
measured. The device selects AC or DC automatically according to
which measured value is larger. The AC or DC symbol displays blinking
in auto detect mode.
To select AC or DC manually, press the
them. The symbol corresponding to the choice will then display.
The measured value is displayed on the screen.
, until the "full screen" display ends and a beep is emitted.
.
(yellow) button down while turning the
(yellow) button to toggle between
21
(yellow) button is