7
Performance Tests and Calibration
Recommended Test Equipment
Application
DC Voltage
DC Current
Resistance
AC Voltage
AC Current
Frequency
Capacitance
Duty Cycle
Nanosiemens
Diode
Frequency
Counter
Temperature
Square Wave
Short
124
The test equipment recommended for the performance
verification and adjustment procedures is listed below. If the
exact instrument is not available, substitute calibration
standards of equivalent accuracy.
A suggested alternative method would be to use the Agilent
3458A 8½ - Digit Digital Multimeter to measure less accurate
yet stable sources. The output value measured from the
source can be entered into the instrument as the target
calibration value.
Table 4
Recommended Test Equipment
Recommended Equipment
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Agilent 33250A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Fluke 5520A
Agilent 33250A
Fluke 5520A
Agilent 53131A and Agilent 34401A
Shorting Plug - Dual banana plug with copper
wire short between 2 terminal
Recommended Accuracy Requirements
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
Agilent U1251A/U1252A User's and Service Guide