Fuse
1
2
Agilent U1251A/U1252A User's and Service Guide
6 Gently remove the defective fuse by prying one end of the
fuse loose and removing it out of the fuse bracket.
7 Replace with a new fuse of the same size and rating.
Make sure the new fuse is centered in the fuse holder.
8 Ensure that the rotary switch on the top case and the
circuit board switch stay on the OFF position.
9 Then re- fasten the circuit board and the bottom cover.
10 Refer to the table below for the part number, rate and
size of the fuses.
Agilent Part Number
2110-1400
2110-1402
Figure 53
Rating
440mA/1000V
11A/1000V
Fuse 2
Fuse replacement
Maintenance
Size
10 mm x 35 mm
Fast blow Fuse
10 mm x 38 mm