Agilent Technologies U1251A User Manual page 150

Handheld digital multimeter
Table of Contents

Advertisement

7
Performance Tests and Calibration
Step
Test Function
5
Press
button to go to the
mV mode
Turn the rotary switch to the W
6
position
150
Cal Range
Input
50mV
3 mV, 1 kHz
30 mV, 1 kHz
30 mV, 50 kHz
500mV
30 mV, 1 kHz
300 mV, 1 kHz
300 mV, 50 kHz
1000mV
30 mV, 1 kHz
1000 mV, 1 kHz
1000 mV, 50 kHz
Short
Dual Banana
Plug with copper
wire short
between 2
terminals
50 MW
Input terminal
open (Remove
any test leads
and Short Plugs
from the input
terminal)
10 MW
5 MW
3 MW
500 kW
300 kW
50 kW
30 kW
3k W
5 kW
500 W
300 W
Agilent U1251A/U1252A User's and Service Guide
Cal Item
U1251A
U1252A
03.000 mV
03.000 mV
30.000 mV
30.000 mV
N/A
30.000 mV
030.00 mV
030.00 mV
300.00 mV
300.00 mV
N/A
300.00 mV
0030.0 mV
0030.0 mV
1000.0 mV
1000.0 mV
N/A
1000.0 mV
SHort
SHort
oPEn
oPEn
10.000 MW
10.000 MW
3.0000 MW
3.0000 MW
300.00 kW
300.00 kW
30.000 kW
30.000 kW
3.0000 kW
3.0000 kW
300.00 W
300.00 W