Performance Tests and Calibration
Step
Test Function
Turn the rotary switch to the W
8
position
9
Press
button to go to ns
mode
10
Turn the rotary switch to the Hz/
position (for model
U1252A), to
model U1251A)
11
Press
button to go to
frequency counter mode
12
Press
frequency counter mode divide by
100
13
Turn the rotary switch to the
TEMP/
position
134
Range
500 W
5 kW
50 kW
500 kW
5 MW
50 MW
500 MW
500 nS
Diode
position (for
999.99 kHz
[6]
button to go to
99.999 MHz
10.000 nF
[7]
100.00 nF
1000.0 nF
10.000 μF
5520A Output
500 W
5 kW
50 kW
500 kW
5 MW
[4]
50 MW
450 MW
[5]
50 nS
1 V
33250A Output
200 mVrms,
100 kHz
600 mVrms,
10 MHz
5520A Output
10.000 nF
100.00 nF
1000.0 nF
10.000 μF
Agilent U1251A/U1252A User's and Service Guide
Error from nominal 1 year
U1251A
U1252A
[3]
± 350 m
± 500 mW
[3]
± 4.5 W
± 3 W
± 45 W
± 30 W
± 450 W
± 300 W
± 10.5 kW
± 8 kW
± 0.510 MW
± 0.505 MW
N/A
± 36.1 MW
± 0.7 nS
± 0.6 nS
± 1 mV
± 1 mV
N/A
± 52 Hz
N/A
± 5.2 kHz
± 0.108 nF
± 0.108 nF
± 1.05 nF
± 1.05 nF
± 10.5 nF
± 10.5 nF
± 0.105 μF
± 0.105 μF
[3]
W