U3402-90001.book Page 63 Friday, July 24, 2009 4:04 PM

Recommended Test Equipment

U3402A User's and Service Guide
Downloaded from
Elcodis.com
electronic components distributor
The test equipments recommended for the performance verification
procedures are listed below. If the exact instrument is not available, substitute
calibration standards of equivalent accuracy.
A suggested alternate method would be to use the Agilent 3458A 8½ - digit
digital multimeter to measure less accurate yet stable sources. The output
value measured from the source can be entered into the instrument as the
target calibration value.

Table 4-1 Recommended test equipments

Application
Recommended equipment
Zero calibration
Shorting plug — Dual banana plug with
copper wire short between the two terminals
DC voltage
Fluke 5520A
DC current
Fluke 5520A
Resistance
Fluke 5520A
AC voltage
Fluke 5520A
AC current
Fluke 5520A
Frequency
Fluke 5520A
Diode
Fluke 5520A
Performance Test
Recommended accuracy
requirements
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
<1/5 instrument 1 year spec
63