2 Making Measurements
119
In the range of 0–500 Ω, the beeper will sound if the
resistance value falls below 10 Ω. For other ranges, the
beeper will sound if the resistance falls below the typical
values indicated in the table below.
Measuring range
500.00 Ω
5.0000 kΩ
50.000 kΩ
500.00 kΩ
5.0000 MΩ
50.000 MΩ
500.00 MΩ
The conductance measurement enables the measurement of
very high resistance of up to 100 GΩ. As the high- resistance
readings are susceptible to noise, you can capture the
average readings by using the Dynamic Recording mode. See
Figure 17 on page 59.
Agilent U1251A/U1252A User's and Service Guide
Beeper sounds when
Ω < 10
Ω < 100
Ω < 1k
Ω < 10k
Ω < 100k
Ω < 1M
Ω < 10M

Hide quick links: